![Cardoline 80509531 Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/cardoline/80509531/80509531_technical-and-service-manual_2556678023.webp)
23
AR1200VIEW
CARDIOLINE
®
is an et medical devices SpA brand
w w w . c a r d i o l i n e . b i z
12 Spare parts list
The code numbers of the spare parts are listed in table 12.1.
The spare part code is also indicated on the label identifying the main subassemblies inside the device.
To order a spare part, use the corresponding code number.
List of spare parts for the ar2100 view device code 80600062
Spare part code
Description
Ref Note
69700788
FUSE 0,5 AT 5X20 MM (10PCS)
Installed on mains plug
69701148
FUSE 5 A SMF SLO-BLO SMD (10PCS)
Installed on mother board
69701307
BLACK MARK PHOTOSENSOR BOARD AR 600/AR 1200
69701352
BATTERY PACK (5+5) X 1,2V
69701354
MOTOR AS GERAS AR 1200 II SERIE
69701358
PAPER S PAPER BOARD GUIDE AR 1200
69701360
MAINS PLUG SOCKET AR 1200
69701361
THERMAL PRINTER – WRITING SYSTEM AR 1200
69701363
MAINS POWER LABEL 115V E 230V
69701440
D GR. SUPPORTO AR 1200 VIEW
69701441
DISPLAY GLASS FOR AR 1200 VIEW
69701492
ELECTRONIC KEYBOARD BOARD AR 1200 VIEW EXT.
69701502
MOTHER BOARD+IRDA AR1200VIEW EXT C/LINE
69701502E
MOTHER BOARD+IRDA AR1200VIEW EXT C/LINE
69701503
KTRADE MARK LABEL AR1200VIEW EXT C/LINE
69701511
BATTERY COVER (WHITE)
69701515
PAPER GUIDE + RUBBER ROLL AR 1200 (WHITE)
69701517
BOTTOM CASE AR1200VIEW (WHITE)
69701578
UPPER CASE AR1200VIEW (WHITE)
Table 12.1
* The replacement motherboard is supplied complete with basic software in English and S/N 00000000.
To reinstall the configuration before the replacement of the motherboard, the following data must be supplied:
device code number (REF)
serial number (SN)
language
options purchased
et medical devices
will send the corresponding firmware (binary file to load through the Loader application), which will install the correct S/N
and enable the options previously purchased.
Warning: Once the firmware has been reloaded with a specific S/N it cannot be further modified.
Note: The data requested can be printed out directly by the instrument, using the info function on the
self-test menu.