background image

8535-M074-O1 Rev C 

 6745 Baby Scale Operation 

9

PRINTER OUTPUT 

The Model 6745 Baby Scale has a RS-232 serial port that may be connected to an optional 
printer to record weight. The output connector is located on the rear panel (see Figure No. 6).  
 
When the weight display is locked on the baby’s weight and the Weight Lock annunciator is 
turned on, the weight data may be transmitted by pressing the 

PRINT

 key or on demand with 

receipt of a command from a computer). 

NOTE:

 If the threshold feature was enabled during 

setup and calibration, with a printer attached to the scale, the weight data will automatically 
print when the weight display locks on the baby’s weight. 

 

  

 

  

 

 
 
 
 

Figure No. 6 

 
The serial port can be configured during the setup and calibration procedure or during the 
setup review operation. Using either method, it is possible to select the baud rate, data format 
and print ticket format. 

NOTE:

 The scale is shipped with the baud rate set to 9600 baud. 

 
 

SERIAL DATA FORMAT 

If the scale is connected to a computer, it will transmit a single set of weight data each time the 
computer sends an ENQ (hex 05) or a SMA weight request (W). Examples and explanation of 
the data format transmitted are shown below.  
 
The host device (computer) sends: 

ENQ (hex 05) or <lf> W <cr>. 

The scale will respond with: 

<lf><s><r><n><m><xxxxxxxxxx><uuu><cr>. 

Where: 

lf =  

Line Feed 

 

s =  

Flags 

Z= center of Zero 
O = Over cap 
E = zero Error, 
e = weight not currently being displayed 

r = 

Range 

1, 2, 3, ... 

n = 

Mode 

G = Gross 

m = 

Motion 

M = Motion, " "(

blank

) = no motion 

xxxxx:xx.x = 
xxxxxx.xxx = 

Weight 

Ten digits (includes decimal point), lb/oz mode 
Ten digits (includes decimal point), kg mode 
Weight is right justified. 

uuu =  

Units 

l/O^ (lb & oz), kg^ (kilograms), (^ = space) 

cr =  

Carriage Return 

(hex 0D) 

PIN NO. 

FUNCTION 

DATA INPUT (RXD)

 

DATA OUTPUT (TXD)

 

SIGNAL GROUND (GND)

 

 

 

 1     

 

 

 

 

 5    

 

 

 
 

 

    6 

 

 

 

    9 

 

 

Содержание 6745

Страница 1: ...n 1 MODEL 6745 BABY SCALE Operation Manual CARDINAL SCALE MFG CO 8535 M074 O1 Rev C PO BOX 151 WEBB CITY MO 64870 Printed in USA 03 15 PH 417 673 4631 FAX 417 673 5001 www detectoscale com Technical Support Ph 866 254 8261 techsupport cardet com ...

Страница 2: ...8535 M074 O1 Rev C 6745 Baby Scale Operation 2 I ...

Страница 3: ...ow Battery Indicator 5 Battery Status 5 Battery Charging 5 Placing the Scale 5 KEYPAD FUNCTIONS 6 ANNUNCIATORS 7 OPERATION 8 Basic Weighing Operation 8 Operating Hints 8 PRINTER OUTPUT 9 SERIAL DATA FORMAT 9 SETUP AND CALIBRATION 10 SETUP REVIEW 16 ERROR AND STATUS DISPLAYS 17 BEFORE YOU CALL SERVICE 17 PART IDENTIFICATION 18 I Serial Number_______________________ Date of Purchase ________________...

Страница 4: ...ca the device should be disposed of in accordance with the local laws regarding the disposal of waste electrical and electronic equipment It is everyone s responsibility to help maintain the environment and to reduce the effects of hazardous substances contained in electrical and electronic equipment on human health Please do your part by making certain that this device is properly disposed of The...

Страница 5: ...cate and remove the battery access cover on the bottom of the scale Install six 6 C size batteries and then replace the cover Step 3 Turn the scale over to the normal operating position Step 4 Press the ON key to turn the scale on If desired place a blanket or pad on the baby tray and press the ZERO key to reset the weight display to zero Step 5 Place the baby on the scale Wait a few seconds for t...

Страница 6: ...port of the scale become necessary Level Adjustment Check to make certain the scale is level The level indicator is located in the center of the weighbridge under the baby tray Remove the baby tray and observe the level bubble see Figure No 1 If the scale is not level the bubble will not be centered loosen the locking nut on all four 4 mounting feet see Figure No 2 and adjust them as required to c...

Страница 7: ...ge Step 3 Slide the baby tray onto the weighbridge until you hear the release knob click Step 4 Verify the baby tray is locked in place by making certain that it cannot be slide to the left or right WARNING Make absolutely certain that the baby tray is locked in place before using the scale Be sure that the locking knob is in the locked position Failure to lock the baby tray in place can result in...

Страница 8: ...he thumb screw retaining it Step 3 Remove the panel lift straight up and slide it out exposing the battery holder Step 4 Install 6 C size batteries in the holder making certain they are positioned in accordance with the polarity markings located in the battery holder Refer to Figure No 4 Step 5 After placing all 6 batteries in the battery holder replace the panel on the bottom of the scale slide t...

Страница 9: ...he scale can still be operated Note that charging the batteries for more than 15 hours will not damage them When the scale is to be turned off pressing the OFF key once will display dashes scrolling across the display indicating the batteries are being charged Pressing the OFF key again will display OFF and turn the scale off If the power supply is disconnected before the 15 hours the scale will c...

Страница 10: ... This key is used to change the weighing units to the alternate units of measurement if selected during setup of the scale WEIGHTING UNITS 2 or 3 For example with pounds displayed lb annunciator turned on pressing this key will change the weighting units to kilograms kg annunciator will turn on NOTE This feature must be enabled during setup and calibration for this key to be operational This key i...

Страница 11: ...t center of the weight display and is turned on when the pounds ounces mode is selected to show that the displayed weight is ounces kg The kg annunciator is located on the right of the weight display and is used to indicate that the displayed units of weight measurement is kilograms LOW BATTERY The low battery annunciator is used with the battery operation It will turn ON to indicate that the batt...

Страница 12: ...nds the Weight Lock annunciator will turn on signaling that the weight display is locked and the baby may be removed from the scale Note that if a printer is attached to the scale the weight data will automatically print Step 4 Remove the baby from the scale Step 5 Record the baby s weight from the weight display Note that after a short period of non use the scale will turn off To resume operation...

Страница 13: ...peration Using either method it is possible to select the baud rate data format and print ticket format NOTE The scale is shipped with the baud rate set to 9600 baud SERIAL DATA FORMAT If the scale is connected to a computer it will transmit a single set of weight data each time the computer sends an ENQ hex 05 or a SMA weight request W Examples and explanation of the data format transmitted are s...

Страница 14: ...nd hold the calibration switch 3 Press the ON key 4 The display will show int The indicator is now ready for setup and calibration Figure No 7 Calibration Access Screw During the setup and calibration process it will be necessary to enter data using the scale s keyboard Pressing the PRINT key will show the current value of a setting Pressing the PRINT key again will save the displayed setting valu...

Страница 15: ...eleration of Gravity simply press the PRINT key and the internal calibration factor will be retained To begin calibration press UNITS to select 1 1 YES then press the PRINT key After pressing the PRINT key the display will change to LOAd Load Calibration Weight The scale will now display LOAd which is a prompt for the entry of the calibration weight value and placement of this amount of test weigh...

Страница 16: ... location Repeat the process until all digits have been entered After all digits have been correctly entered press the PRINT key to store the value and advance to the next step Consult the factory for the Acceleration of Gravity value for your location Acceleration of Gravity Operation Location The display will change to show OP9C Press the PRINT key to show the current setting This is the acceler...

Страница 17: ...ering are available They are as follows 0 NO FILTERING 1 MINIMAL FILTERING 2 MODERATE FILTERING 3 CUSTOM FILTERING NOTE Selection 3 Custom Filtering is used when 0 1 or 2 are inadequate F Filter Level If you select Custom Filtering the scale will display F Press the PRINT key to show the current setting for the Filter Level The filter level is a number from 1 to 16 that corresponds to the level of...

Страница 18: ...splayed is acceptable press the PRINT key again to save it Otherwise use the UNITS and ZERO keys to enter a new value and then press the PRINT key to save it Allowable values are 0 through 99 Note that a setting of 0 will disable the threshold lock Power Up Zero With the display showing PUO press the PRINT key to show current setting If the value displayed is acceptable press the PRINT key again t...

Страница 19: ...ave it Otherwise use the UNITS key to select the new setting and then press the PRINT key to save it NOTE The FACTORY setting is 3 9600 baud and may be revised without having to enter the calibration mode The following baud rates are available 0 1200 1 2400 2 4800 3 9600 4 19 200 5 38 400 Serial Data Format With the display showing PrtY press the PRINT key to show the current setting If acceptable...

Страница 20: ... to enter the setup and calibration mode The parameters in the setup review will be processed in the following sequence NiCAd 0 Disable battery charging use Alkaline batteries 1 Enable battery charging use Ni Cad or NiMH 2 Enable battery charging and Force charging of discharged Ni Cad or NiMH batteries tHrSh Enable or Disable the threshold lock and select the weight that the load on the scale mus...

Страница 21: ...Err NOVRAM failure Consult the scale service representative trL Indicates an attempt to zero a weight outside the scale zero range battry Indicates the remaining battery voltage expressed as a percentage of the total battery voltage Err General error invalid keypad entry was attempted E Unit Displayed when the UNITS key is pressed and only a single weighing units mode 0 pound ounces or 1 kilograms...

Страница 22: ...8535 M074 O1 Rev C 6745 Baby Scale Operation 18 PART IDENTIFICATION ...

Страница 23: ...8535 M074 O1 Rev C 6745 Baby Scale Operation 19 PART IDENTIFICATION CONT ...

Страница 24: ...LEVELER SCREW 1 4 20 X 1 S S 10 2 6610 2000 JACK SOCKET 11 2 6610 5007 CABLE CLIP 12 1 6610 5119 BATTERY HOLDER 6 C CELLS 13 1 6650 0018 GASKET MATERIAL 1 X 1 2 X 5 1 4 14 4 6680 0004 WASHER LOCK INT TOOTH 6 Z P 15 2 6680 0052 WASHER LOCK 4 Z P 16 4 6680 0214 RIVET POP BH AL 5 32 DIA X 0 126 187 GR 17 1 8526 B214 08 LOAD CELL SPACER 18 1 8535 B050 0A POWER CABLE 19 1 8535 B051 0A CABLE BATTERY 20 ...

Страница 25: ...8535 M074 O1 Rev C 6745 Baby Scale Operation 21 PART IDENTIFICATION CONT ...

Страница 26: ...8535 M074 O1 Rev C 6745 Baby Scale Operation 22 PART IDENTIFICATION CONT ...

Страница 27: ...21 1554 SCW SOCKET HD CAP SCW 025 20 1 50D 7 2 6024 0039 WASHER LOCK HELICAL 1 4 REG 8 5 6024 1010 WASHER LOCK INT TOOTH 10 Z PL 9 1 6560 1072 ADHESIVE LOCTITE 10 1 6650 0087 LABEL MADE IN THE USA 11 8 6680 0173 POP RIVET 12 1 6690 0001 BUBBLE LEVEL 13 2 8535 B060 08 BABY TRAY RECEIVER 14 1 8535 B062 08 LOAD CELL SPACER 15 1 8535 B065 08 WARNING LABEL BABY SCALE 16 1 8535 B067 0A WEIGHBRIDGE WELDM...

Страница 28: ...8535 M074 O1 Rev C 6745 Baby Scale Operation 24 ...

Отзывы: