12. Maintenance
WARNING:
Risk of Shock. Risk of Shock. Disconnect equipment from main power before attempting any maintenance to
equipment or its controls unless stated otherwise.
12.1 Cleaning the Cabinet Interior
To clean the cabinet interior, remove the shelves, drawers or baskets following the instructions in
Section 7.3, Section 7.4
and
Section 7.5
.
Use a solution of water and a mild detergent for cleaning. Rinse the interior storage components and wipe them dry with a soft cloth.
12.2 Cleaning the Condenser Filter
Clean the condenser filter every three months. There is one condenser filter located in the back cage of the unit that can be accessed
without removing the back cage or disconnecting the power.
To clean the filter, complete the following steps:
1. Remove the filter by pulling upwards through the slot in the back cage.
2. Shake the filter to remove loose dust.
3. Rinse the filter in clean water.
4. Shake the excess water from the filter and let it dry.
5. Reinstall filter by pushing downwards through the slot in the back cage.
CAUTION:
Risk of Abrasion. Do not pull the filter downwards from the bottom. The condenser has sharp surfaces.
12.3 Cleaning the Condenser
Condensers should be cleaned at least every six months; more often if the laboratory area is dusty. In heavy traffic areas,
condensers load with dirt more quickly. Failure to keep the condenser clean can result in equipment warm-up or erratic
temperatures.
CAUTION:
Risk of Abrasion. Never clean around the condensers with your fingers. Some surfaces are sharp.
The condenser is located in the top rear of the machine compartment. To clean the condenser, complete the following steps:
1. Disconnect the power.
2. Remove the back filter.
3. Vacuum the condenser and clean up any loose dust.
4. Replace the back filter.
5. Reconnect power.