CARDIAC SCIENCE AEDs
INSTRUCTIONS FOR USE
12
IMMUNITY (CON’T)
Field
Standard or Compliance
Magnetic
IEC 61000-4-8 (2001)
IEC 60601-2-4 (2002), Section 36.202.8
AAMI DF39, Section 3.3.21.2.3 80A/m, 47.5Hz – 1,320Hz
ESD
IEC 61000-4-2, Level 3
IEC 60601-2-4 (2002), Section 36.202.2
6KV contact discharge, 8KV air gap discharge
ENVIRONMENTAL CONDITIONS
Condition
Standard or Compliance
Free Fall Drop
IEC 60068-2-32 (1975) AM 2 (1990), 1 meter
Bump
IEC 60068-2-29 (1987), 40g and 6000 bumps
Vibration
(Random)
IEC 60068-2-64 (1993): 10Hz – 2KHz, 0.005 – 0.0012 g
2
/Hz
Vibration
(Sine)
IEC 60068-2-6 (1995): 10Hz – 60Hz, 0.15 mm and
60Hz – 150Hz, 2g
Enclosure
Protection
IEC 60529 (2001), IP24
SHIPPING AND TRANSPORTATION CONDITIONS
ISTA Procedure 2A
Содержание FirstSave AED 3g
Страница 27: ...P N 112 0036 401 Rev A...