70-01047-01 B
Maintenance
33
Testing the battery
!
WARNING! Fire or explosion.
Never remove the battery pack and attempt to recharge using an
external battery charger.
The ECG device automatically monitors battery status. No
additional battery tests or calibration are required.
If the battery does not retain a charge for more than 30 minutes
of operation, see the
User’s Guide
.
Annually
A complete safety and component check must be performed
annually by a qualified service technician.
The testing must include a 1 mV calibration pulse test to verify
the gain hardware is properly calibrated.
As needed
Perform these functions as necessary.
To...
Do this...
Replace the power
cord
1.
Disconnect the AC power cable from the
wall socket.
2.
Disconnect the external power supply (brick
cable) from the back of the ECG device.
3.
Replace the portions of the cable as
necessary and follow the instructions
Replace the
patient cable
1.
Disconnect the patient cable from the ECG
device.
2.
Replace the patient cable with the Cardiac
Science-approved cable, see the
Accessory
List
.