70-01047-01 B
Warnings and cautions
11
!
WARNING! False readings
The lead status check does not detect lead reversals.
Always ensure the correct lead placement is used for the lead
configuration selected.
!
WARNING! Misdiagnosis
Dirty, compromised, or unprepared skin may cause incorrect
readings.
Before applying electrodes, ensure the skin is clean, unbroken,
and properly prepared.
!
WARNING! Infection
Lead placement on compromised skin may cause infection.
Before applying electrodes, ensure the skin is clean, unbroken and
properly prepared.
!
Caution: Patient skin irritation
When applying electrodes, skin preparation, electrolyte solution,
or electrode material may cause skin irritation. Monitor the
electrode site and, if irritation occurs, use alternative electrolyte or
electrodes.
!
Caution: Patient safety
Disposable electrodes are single use only. Never re-use disposable
electrodes.
!
Caution: Baseline drift
Never mix patient electrode types or brands. Dissimilar metals or
other incompatibilities may cause considerable baseline drift and
may increase trace recovery time after defibrillation.
Defibrillation safety
!
WARNING! Patient cable safety
To ensure defibrillation protection, you must use the Cardiac
Science-approved patient cable. Follow all instructions for the
patient cable.