AINTENANCE

 

4.1 

General Maintenance 

 

No routine maintenance is required other than removal of soot deposits mentioned in 3.2 and the 
occasional replacement of consumable items. 

 

The furnace outer surface may be cleaned with a damp cloth. Do not allow water to enter the 
interior of the case or chamber. Do not clean with organic solvents. 

4.2 

Calibration 

 

After prolonged use the controller and/or thermocouple could require recalibration. This would be 
important for processes which require accurate temperature readings or which use the furnace 
close to its maximum temperature. A quick check using an independent thermocouple and 
temperature indicator should be made from time to time to determine whether full calibration is 
required. Carbolite can supply these items. 

 

Depending on the controller, the controller manual may contain calibration instructions. 

4.3 

After Sales Service 

Carbolite’s service division (Thermal Engineering Services) has a team of Service Engineers 
capable of repair, calibration and preventive maintenance of furnace and oven products at our 
customers’ premises throughout the world. We also sell spares by mail order. A telephone call or 
fax often enables a fault to be diagnosed and the necessary spare part despatched.  

 

Each furnace has its own record card at Carbolite. In all correspondence please quote the serial 
number, model type and voltage given on the rating label of the furnace. The serial number and 
model type are also given on the front of this booklet when supplied with a furnace. 

 

To contact Thermal Engineering Services or Carbolite see the back page of this manual. 

4.4  Recommended Spares Kits 

 

Carbolite can supply individual spares, or a kit of the items most likely to be required. Ordering a 
kit in advance can save time in the event of a breakdown. Each kit comprises one thermocouple, 
one solid state relay, one door insulation piece, and set of elements. 

 

When ordering spares please quote the model details as requested above. 

4.5 

Power Adjustment 

 

The furnace control system incorporates electronic power limiting, but in these models the power 
limit is set to 100% (but see section 8.2 for any exceptions). The power limit parameter 

OP.Hi

 may 

be accessible to the operator, but should not generally be altered. 

 

Occasionally the power limit is set to zero to permit demonstration of the controls without the 
heating elements taking power. In this case the power limit may be reset to its standard value to 
re-enable heating. 

5.0 

 

Содержание HRF 7/22

Страница 1: ...to the electricity supply CONTENTS Section page 1 0 Symbols Warnings 1 2 0 Installation 3 3 0 Operation 5 4 0 Maintenance 6 5 0 Repairs Replacements 7 6 0 Fault Analysis 9 7 0 Circuit Diagrams 10 8 0...

Страница 2: ...glows Heat Light the adjacent light glows or flashes to indicate that power is being supplied to the elements 1 2 Warning Symbols DANGER of electrical shock read any warning printed by this symbol DA...

Страница 3: ...re that there is free space around the furnace Do not obstruct any of the vents in the control section they are needed to keep the controls cool Ensure that the furnace is placed in such a way that it...

Страница 4: ...ses customer fusing is preferred but not essential Furnace with supply cable either wire directly to an isolator or fitted with a line plug Furnace without supply cable a permanent connection to a fus...

Страница 5: ...ating elements are isolated Find and correct the cause before resetting the overtemperature controller according the instructions supplied To switch the furnace off set the Instrument switch to O If t...

Страница 6: ...by mail order A telephone call or fax often enables a fault to be diagnosed and the necessary spare part despatched Each furnace has its own record card at Carbolite In all correspondence please quote...

Страница 7: ...mend reference to the European Ceramic Fibre Industry Association Bulletin No 11 and the UK Health and Safety Executive Guidance Note EH46 We can provide further information on request Alternatively o...

Страница 8: ...ng to the new element Remove the chimney open the furnace door and slide out the chamber liner Withdraw the element through the front of the furnace Insert the new element through the glands and refit...

Страница 9: ...no lights glowing on the controller The SUPPLY light is ON The controller may be faulty or not receiving a supply due to a faulty switch or a wiring fault The SUPPLY light is OFF Check the supply fuse...

Страница 10: ...ply light Instrument Switch E coil temperature controller thermocouple overtemp controller thermocouple L N F1 F2 Filter if fitted contactor R BU Fan Motor Capacitor BR BU Fan Motor Capacitor BR BU he...

Страница 11: ...fuses Required if no supply cable fitted Recommended if cable fitted See rating label for amperage see table below for fuse rating Model phases Volts Supply Fuse Rating HRF 7 22 1 phase 220 240 15A or...

Страница 12: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Отзывы: