background image

 

HRF 

 

 
MF05 

O

PERATION

 

 

The instructions for operating the temperature controller are given in a separate manual.   

 

If the furnace is fitted with a time switch, see also the supplementary manual MS03. 

 

If cascade control is fitted, see the supplementary manual MS07. 

3.1 

Operating Cycle 

 

The furnace is fitted with a combined Supply light and Instrument switch. The light is on 
whenever the furnace is connected to the supply. The switch cuts off power to the control circuit. 
The Instrument switch does not cut power to the fan: the fan is on when furnace is connected to 
the supply. 

 

Connect the furnace to the electrical supply. The Supply light should glow. 

 

Operate the instrument switch to activate the temperature controller; the 

O

 position is 

off

, the 

I

 position 

on

. The controller becomes illuminated and goes through a short test cycle. 

 

Adjust the temperature – see the controller manual. 

 

Overtemperature option only.

 If the overtemperature controller has not yet been set as required, 

set it and activate it according to the instructions in the appropriate manual. 

 

Unless a time switch is fitted and is off, the furnace starts to heat up. The Heat light glows steadily 
at first and then flashes as the furnace approaches the desired temperature or a program setpoint. 

 

Overtemperature option only.

 If the overtemperature trip operates then an indicator in the 

overtemperature controller flashes, and the heating elements are isolated. Find and correct the 
cause before resetting the overtemperature controller according the instructions supplied. 

 

To switch the furnace off, set the Instrument switch to 

O

. If the furnace is to be left off, isolate it 

the electrical supply. 

3.2 

General Operating Advice 

 

Heating element life is shortened by use at temperatures close to maximum. Do not leave the 
furnace at high temperature when not required. The maximum temperature is shown on the 
furnace rating label and on the back page of this manual. 

 

When heating materials which produce smoke or fumes, the chimney must be correctly fitted and 
unobstructed. Otherwise, soot can accumulate in the chamber and could possibly cause an 
electrical breakdown of the heating element.  

 

If the furnace is used to heat materials which emit smoke or fumes, regularly heat it up to 
maximum temperature for one hour without load to burn away the soot.  

3.3 

Atmospheres 

 

When an optional gas inlet is fitted there is a label near the inlet saying "INERT GAS ONLY". In 
practice 

inert

 or 

oxidising

 gases may be used, but 

not combustible or toxic gases

. Chamber 

furnaces are not gas tight, so it should be understood that gas usage may be high, and that the 
chamber is likely always to contain some air. Residual oxygen levels of 1% are to be expected. 

3.4 

Operator Safety 

 

The furnace has mineral insulated heating elements, and does not require a door safety switch. 

 

Avoid burns. Carbolite can supply tongs, face masks, and heat resistant gloves. Before you 
remove a hot object from the furnace make sure you have a safe place to put it down. 

4.0 

 

Содержание HRF 7/22

Страница 1: ...to the electricity supply CONTENTS Section page 1 0 Symbols Warnings 1 2 0 Installation 3 3 0 Operation 5 4 0 Maintenance 6 5 0 Repairs Replacements 7 6 0 Fault Analysis 9 7 0 Circuit Diagrams 10 8 0...

Страница 2: ...glows Heat Light the adjacent light glows or flashes to indicate that power is being supplied to the elements 1 2 Warning Symbols DANGER of electrical shock read any warning printed by this symbol DA...

Страница 3: ...re that there is free space around the furnace Do not obstruct any of the vents in the control section they are needed to keep the controls cool Ensure that the furnace is placed in such a way that it...

Страница 4: ...ses customer fusing is preferred but not essential Furnace with supply cable either wire directly to an isolator or fitted with a line plug Furnace without supply cable a permanent connection to a fus...

Страница 5: ...ating elements are isolated Find and correct the cause before resetting the overtemperature controller according the instructions supplied To switch the furnace off set the Instrument switch to O If t...

Страница 6: ...by mail order A telephone call or fax often enables a fault to be diagnosed and the necessary spare part despatched Each furnace has its own record card at Carbolite In all correspondence please quote...

Страница 7: ...mend reference to the European Ceramic Fibre Industry Association Bulletin No 11 and the UK Health and Safety Executive Guidance Note EH46 We can provide further information on request Alternatively o...

Страница 8: ...ng to the new element Remove the chimney open the furnace door and slide out the chamber liner Withdraw the element through the front of the furnace Insert the new element through the glands and refit...

Страница 9: ...no lights glowing on the controller The SUPPLY light is ON The controller may be faulty or not receiving a supply due to a faulty switch or a wiring fault The SUPPLY light is OFF Check the supply fuse...

Страница 10: ...ply light Instrument Switch E coil temperature controller thermocouple overtemp controller thermocouple L N F1 F2 Filter if fitted contactor R BU Fan Motor Capacitor BR BU Fan Motor Capacitor BR BU he...

Страница 11: ...fuses Required if no supply cable fitted Recommended if cable fitted See rating label for amperage see table below for fuse rating Model phases Volts Supply Fuse Rating HRF 7 22 1 phase 220 240 15A or...

Страница 12: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Отзывы: