ELF (B) 

 

 

R

EPAIRS 

& R

EPLACEMENTS

 

7.1  Safety Warning – Disconnection from Supply 

 

Always ensure that the furnace is disconnected from the supply before repair work 
is carried out. 

7.2 

Safety Warning - Refractory Fibrous Insulation 

 This furnace contains refractory fibres in its thermal insulation. These materials may 

be in the form of fibre blanket or felt, vacuum formed board or shapes, mineral wool 
slab or loose fill fibre.  

 

Normal use of the furnace does not result in any significant level of airborne dust from these 
materials, but much higher levels may be encountered during maintenance or repair. 

 

Whilst there is no evidence of any long term health hazards, we strongly recommend that safety 
precautions are taken whenever the materials are handled. 

Exposure to dust from fibre that has been used at high temperatures may cause 
respiratory disease. 

When handling fibre always use an approved mask, eye protection, gloves and long 
sleeved clothing. 

Avoid breaking up waste material. Dispose of waste fibre in sealed containers. 

After handling rinse exposed skin with water before washing gently with soap (not 
detergent). Wash work clothing separately. 

 

Before commencing any major repairs we recommend reference to the European Ceramic Fibre 
Industry Association Bulletin No. 11 and the UK Health and Safety Executive Guidance Note 
EH46. 

 

We can provide further information on request. Alternatively our service division can quote for 
any repairs to be carried out at your premises or ours. 

7.3 

Temperature Controller Replacement 

 

The 201 controller is fitted inside the furnace base, which can be separated from the top by 
removal of six screws. 

 

Before handling the controller: 

wear an anti-static wrist strap

 or otherwise avoid any possibility 

of damage to the unit by static electricity.  

 

See the instructions supplied with the replacement controller. 

7.4  Solid-state Relay Replacement 

 

Disconnect the furnace from the supply and remove the furnace back panel. 

 

Make a note of the wire connections to the solid state relay, and disconnect them. 

 

Remove the solid state relay from the base panel. 

 

Replace and reconnect the solid state relay ensuring that the heat-conducting thermal pad is 
sandwiched between the relay and the base panel or aluminium plate. Alternatively a thin layer of 
white, heat-conducting silicon paste may be applied between the new relay and the base panel or 
aluminium plate. 

 

Replace the removed panel. 

7.5 

MF01 

15 

Содержание ELF 11/14B

Страница 1: ...ty supply CONTENTS Section page 1 0 Introduction 1 2 0 Installation 2 3 0 Operation 5 4 0 Controller Operation 7 5 0 Overtemperature Control Option 13 6 0 Maintenance 14 7 0 Repairs Replacements 14 8...

Страница 2: ...ns The B versions may be set to operate on any single phase voltage in the range 200 240V or 100 120V to order by correct setting of the power limit parameter in the controller Because of model change...

Страница 3: ...pply 2 3 Chimney The chimney is a short length of ceramic tubing If it is supplied unfitted then fit it through the hole in the top of the furnace case If the furnace is to be used to heat substances...

Страница 4: ...the supply cable directly to an isolator or fit it with a line plug The plug or isolator should be within easy reach of the operator to permit quick disconnection of the power For operator safety the...

Страница 5: ...controller setpoint to a temperature close to the desired maximum Materials such as case hardening compounds and other reactive salts may attack the wire elements causing premature failure The insula...

Страница 6: ...ELF B 4 0 6 MF01...

Страница 7: ...oint ramp rate is Off A single press of the scroll key 4 displays the temperature units further presses reveal the parameters in the current list indicated in the Navigation Diagram To return to the H...

Страница 8: ...ate the possible difference between running without and with a ramp to setpoint value depending on the load and the value used 4 6 Navigation Diagram OP Hi SPrr m A w SP Home List Input List Output Li...

Страница 9: ...he timer starts timing immediately At the end of the timing period control switches off i e goes into Manual to allow cooling and End flashes on the display On during the timing period Off from the en...

Страница 10: ...r changes to run Note that the tmr display shows 0 zero during the last minute of timing and also shows 0 when the time has expired The timer light indicates whether timing is still in progress Method...

Страница 11: ...rs or other system errors User calibration allows compensation for such errors and the 2132 allows for a user 2 point calibration This setting is password protected to avoid accidental alteration Page...

Страница 12: ...once to display the value of OP Hi it is good practice to write down the original value To alter the parameter to the desired value use Down T or Up S Caution Do not increase the power limit value to...

Страница 13: ...ons the letters are described as m and w With the key pressed the overtemperature setpoint is normally displayed To alter the value use Down and Up To view the measured temperature press Scroll until...

Страница 14: ...on Thermal Engineering Services has a team of Service Engineers capable of repair calibration and preventive maintenance of furnace and oven products at our customers premises throughout the world We...

Страница 15: ...nt Wash work clothing separately Before commencing any major repairs we recommend reference to the European Ceramic Fibre Industry Association Bulletin No 11 and the UK Health and Safety Executive Gui...

Страница 16: ...a socket spanner undo the four bolts that fasten the insulation box These are located under the insulation box in the two cross supports Remove the insulation box by carefully sliding it out through...

Страница 17: ...rder DOOR DOOR PLUG CARRIER DOOR PLUG 7 8 Air Inlet Adjustment The air inlet is on the inner door panel door plug carrier and is supplied in the closed position It can be altered to the open position...

Страница 18: ...visable for an electrician to check the internal circuits Replace any failed fuses with the correct type For safety reasons do not fit larger capacity fuses without first consulting Carbolite The fuse...

Страница 19: ...hts glowing on the controller The SUPPLY light is ON The controller may be faulty or not receiving a supply due to a faulty switch or a wiring fault The SUPPLY light is OFF Check the supply fuses and...

Страница 20: ...n t thermocouple EMC filters circuit boards 9 3 Fuses F1 F3 Refer to the circuit diagram F1 Internal supply fuses 32mm x 6mm type F fitted on EMC filter circuit board s F2 Auxiliary circuit fuses 20mm...

Отзывы: