Effects of Impurities
An Ionization chamber itself does not have intrinsic energy- discrimination capability.
The presence of radioisotope impurities will affect the reading of the instrument
unless the effect of impurities is eliminated by photon filtration as is done with Mo99
breakthrough in Tc99m. However, the presence of low-level radionuclide impurity
does not negate the usefulness of a radioisotope calibrator, if the user is aware of its
presence and has an independently determined calibration including photons arising
from the impurities.
A1 - 8
APPENDIX I
January 08