INITIAL TURN ON PROCEDURES
1. Be sure the interconnecting cable from the Chamber assembly is properly plugged
into the back of the Readout unit.
2. Confirm the power requirements of the instrument.
3. Be sure the power switch located at the rear of the unit is off.
CAUTION
Accidental connection of the power plug into a DC line or to an AC line that exceeds
the specified voltage may result in damage to the instrument's circuits.
4. Plug the power cords into a grounded three-wire outlet of the specified power line.
5. Turn on the Readout using the power switch located at the rear of the unit.
6. Turn on the Printer or computer system.
7. The unit should now display numbers on the front panel.
Note: If numbers are not visible on the display, verify an Isotope push-button and a
range button is selected.
8. With
an
optional printer
connected, press the
button on the front panel and
verify one line of “
measurement data
” is printed on the installed printer.
9. With
a
computer connected
, start the communication program that is loaded on the
computer.
a. Select the “VT-100”, “HyperTerminal”, or other “generic” terminal emulation
program.
b. Use the “SETUP” menu to select the COM Port (1 or 2), 9600 Baud Rate, 1
Stop Bit, NO Parity Bit, and 8 Data Bits.
c. Start the terminal emulation program on the computer.
d. Set the “CAPS LOCK” key on the computer keyboard.
ONLY CAPITAL
LETTERS WILL BE RECOGNIZED
e. From the computer keyboard, type the letter
C
and verify one line of
“
measurement data
” is displayed on the computer monitor.
If the communications program or the printer fails to operate, review the installation
procedure in the Owners manual for the communication program or the printer.
If the problem persists, contact Capintec’s Authorized Service Center for assistance.
ACCEPTANCE TESTING
For the initial Quality Assurance Acceptance Test, all of the quality assurance tests specified
in CHAPTER 5: QUALITY ASSURANCE AND ACCEPTANCE TESTING; SECTION:
ACCEPTANCE TESTING
,
MUST
be performed before the CRC
®
-712MX can be used in
normal operation.
4 - 4
SYSTEM SETUP
January 08
Содержание CRC-712MX
Страница 2: ......
Страница 19: ...CAPINTEC INC CRC 712MX Figure 3 1 CRC 712MX Front Panel January 08 GENERAL OPERATING INSTRUCTIONS 3 3 ...
Страница 21: ...CAPINTEC INC CRC 712MX Figure 3 2 CRC 712MX Back Panel January 08 GENERAL OPERATING INSTRUCTIONS 3 5 ...
Страница 22: ......
Страница 29: ...CAPINTEC INC CRC 712MX Figure 4 3 January 08 SYSTEM SETUP 4 7 ...
Страница 30: ......
Страница 36: ......
Страница 48: ......
Страница 62: ......
Страница 64: ......
Страница 68: ......