CAPINTEC, INC.
CRC
®
-712MX
Performance
Activity Range
There are six standard ranges whose full-scale values are as follows:
Curies
Becquerels
1.999 mCi
.1999 GBq
19.99 mCi
1.999 GBq
199.9 mCi
19.99 GBq
1.999 Ci
199.9 GBq
19.99 Ci
1.999 TBq
799.9 Ci
19.99 TBq
Digital Readout
The activity measurement is displayed on a 4-digit, seven-segment (20mm. height)
LED display with a floating decimal point. Measurement over-range is set for 800.0Ci
(20.00TBq) and is indicated by blanking of the display. Measurements above 200.0Ci
(0.800TBq) have reduced accuracy and are indicated by display flashing. The analog
to digital converter has an inherent uncertainty of, ±1 count on the least significant
digit. The units of measurement are displayed as back-lighted measurement
characters after the numeric display.
Iometer (Low Input Impedance Electrometer)
Accuracy:
±3% of reading.
Response time:
90% of full scale reading within 2 seconds for activity
measurements of 20 millicuries (2000GBq) or more.
Noise: RMS
measurement
fluctuations:
2 counts in the last digit for the top 3 Ci (TBq) ranges
6 counts in the last digit for 20mCi (2GBq) & 200mCi (20GBq)
ranges
30 counts in the last digits of the 2.000mCi (.2GBq) range
Detector
Detector Linearity:
chamber saturation 2% for 20Ci (740GBq) of Tc99m.
Detector Response: 25keV to 3MeV.
±2% of mean for radioisotopes with major photon radiation of
over 0.1MeV (when chamber response is normalized for Co60
and Co57 radiations).
January 08
FUCTIONAL & TECHNICAL DESCRIPTION
2 - 5
Содержание CRC-712MX
Страница 2: ......
Страница 19: ...CAPINTEC INC CRC 712MX Figure 3 1 CRC 712MX Front Panel January 08 GENERAL OPERATING INSTRUCTIONS 3 3 ...
Страница 21: ...CAPINTEC INC CRC 712MX Figure 3 2 CRC 712MX Back Panel January 08 GENERAL OPERATING INSTRUCTIONS 3 5 ...
Страница 22: ......
Страница 29: ...CAPINTEC INC CRC 712MX Figure 4 3 January 08 SYSTEM SETUP 4 7 ...
Страница 30: ......
Страница 36: ......
Страница 48: ......
Страница 62: ......
Страница 64: ......
Страница 68: ......