For multiplication, the number can only be multiplied by 10 or 100. For division, the number
can only be divided by 2. Refer to Table 6-1 Calibration Number Limits Table on page 6-16.
To input a new Calibration Number, change a current Calibration Number or restore the
original Calibration Number:
Press
MENU
from the Measurement Screen.
Select
SETUP
. The
Setup Menu
appears.
Select
OTHER
. Input the password (last 3 digits of Readout serial number) and press
ENTER
. Figure 6-1 Other Menu appears.
From the
Other Menu
, select
NUCLIDES
. Figure 6-24 Nuclides Menu will appear.
From the
Nuclide Menu
, select
NEW CAL NUM
. Figure 6-32 Choose Nuclide for Calibration
Number screen will appear.
Cal Numbers
CHOOSE NUCLIDE
Press NUCL or
Pre-set nuclide key
or User key
Figure 6-32 Choose Nuclide for Calibration Number
The nuclide can be chosen via the
NUCL
key, Pre-Set Nuclide key or
USER
key.
Entering New Calibration Numbers
If a the selected nuclide does not have a Calibration Number, Figure 6-33 Change Original
Cal#? Screen will appear.
Sr89
Orig Cal #: NONE
Change Cal #?
Figure 6-33 Change Original Cal#?
March 15
CHAMBER INITIALIZATION
6 - 19
Содержание CRC-25R
Страница 2: ......
Страница 12: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 26: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 58: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 104: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 122: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 192: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 208: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 226: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 227: ......
Страница 228: ......