background image

10. Place outflow tubing in sink

11. Using the front panel arrow keys, set the timer for 30 seconds

12. Press “Start/Stop” button to begin fluid flow

13. If SCOPE BUDDY

®

 Endoscope Flushing Aid is to be left unused for an extended period of time (i.e. 

overnight, weekend, holiday period) please hang the SCOPE BUDDY tubing over taps or an area to 
allow any fluid to drain out of the tubes. 

14. Document Daily Decontamination Procedure on “SCOPE BUDDY Daily Quality Assurance Log”.

Perform the Flow Validation Test by:

1.  Connect Fluid Intake Line to ‘IN’ port and place filter screen in clean water.

2.  Connect Universal Cleaning Adapter Extension Line to ‘OUT’ port, and Flow Verification Tube to 

remaining luer lock end.

3.  Using the front panel arrow keys, set the timer for 30 seconds.

4.  Prime the tubing by pressing the “Start/Stop” button which initiates water flow.

5.  When a steady column of water exits the Flow Verification Tube, press “Start/Stop” to halt flow.

6.  Place open end of pre-primed Flow Verification Tube into empty graduated cylinder.

7.  Using the front panel arrow keys, set the timer for ten (10) seconds.

8.  Press “Start/Stop” button to begin fluid flow into graduated cylinder.

9.  Upon time-out, the display flashes zeros and an audible tone is heard.

10. Remove Flow Verification Tube from graduated cylinder and read dispensed volume.

a.   If volume is ≥ 100ml’s, mark “Pass” on “SCOPE BUDDY Daily Quality Assurance Log”.

SCOPE BUDDY passes the test and is ready for use.

b.   If volume is < 100ml’s, mark “Fail” on “SCOPE BUDDY Daily Quality Assurance Log”.

Do NOT use SCOPE BUDDY and report failure to Medivators Customer Support.

If you have any questions related to this bulletin, please contact your local Medivators technical services 
department or in the USA call 1-800 444 4729.

 

Product Bulletin

50098-974-EN Rev A