Product Bulletin

Medivators Inc. 
14605 28th Avenue North
Minneapolis, MN
55447-4822 USA
Tel: +1.800.444.4729
Fax:  +763.553.3387

www.medivators.com

Medivators BV
Sourethweg 11
6422PC Heerlen
The Netherlands
Tel:  +31.45.5.471.471
Fax: +31.45.5.429.595

Cantel Medical Asia/Pacific Pte. Ltd.
1 A International Business Park
#05-01 Singapore 609933
Tel:   +65.6227.9698
Fax: +65.6225.6848

Cantel Medical Devices (China) Co. Ltd.
Unit 804-805, Innov Tower Block A,
Hongmei Road, Xuhui 200233 Shanghai
Tel:  +86 21 60161380
Fax:  +86 21 61210913

50098-974-EN Rev A

For Distribution To:

•  International Distributors
•  International Management
•  U.S. Field Sales & Clinical Specialists
•  U.S. Technical & Field Service

Date: 

November 2, 2015

RE:

  

SCOPE BUDDY

®

 Endoscope Flushing Aid

 

Daily Quality Assurance 

This bulletin provides alternative instructions for decontamination using chlorine tablets in markets where 
these products are available.

The Daily Decontamination Procedure must be performed at the start of the day before use on any 
endoscopes. (Refer to User Manual for more detailed instructions)

To ensure correct SCOPE BUDDY Endoscope Flushing Aid operation, the Quality Assurance Flow Validation 
Test must be performed prior to the first use of the day.

Contact your local Medivators Customer Support for further assistance.

Prior to First Use of the Day

Perform Daily Decontamination Procedure by:

1.  Locate Fluid Intake Line and place filter screen in decontamination solution in jug (Milton/Peracetic 

Acid/Chlorine Tablets) Use concentration as per manufacturers guidelines.

2.  Locate open end of outflow tubing and place in the decontamination solution which has been made up 

in an empty jug/container (Milton/Peracetic Acid/Chlorine Tablets). 

3.  Using the front panel arrow keys, set the timer for 30 seconds.  

4.  Press “Start/Stop” button to begin fluid flow.

5.  Prime the tubing until a steady column of decontamination solution exits the outflow tubing.

6.  Press “Start/Stop” button to halt fluid flow.

7.  Allow decontamination chemical to remain within Scope Buddy and tubing as per chemical manufacturer’s 

labelling for the appropriate contact time required to achieve decontamination. 

8.  Remove outflow tubing from jug and press start to remove decontamination solution from tubing by 

purging lines with air.

9.  Fill jug/container with clean rinse water and place intake line in jug.

MEDIVATORS

®

 and SCOPE BUDDY

®

 are registered trademarks of Medivators Inc.