OMNERA®
1000-095-071-EN Rev. 1.1
Operator´s Manual
Page 72
2020-09-14
3.2.7.4 Table Flexible Mode
General Description
The
Table flexible mode
is equal to
Auto positioning mode
with functionality added for track-
ing the height of the Table (compare with Wall flexible mode). The mode is intended for Table
examinations.
The stand will upon activation of the servo button, move to the programmed
Table mode
po-
sition and start tracking the Table height, thereby keeping the film focus distance constant.
The
Film focus distance
shown in the display is the actual distance to the detector. The Table
position in the room is set during the installation of the System.
Exposure Validation
Exposure is allowed if the stand is not moving, operating properly (not in an error state) and
the servo button is activated. Movement is allowed in X and alpha direction.
3.2.7.5 Film Tracking Mode
General Description
The
Film Tracking mode
functions as the
Table mode
with added functionality for controlling
the position of the detector in one direction. The mode is intended for fast and easy position-
ing with the X-ray tube always aimed at the center of the detector.
The motorized detector holder will move the detector to the right position. The stand will upon
activation of the servo button move to the pre-programmed
Film-tracking position
and start
tracking the Table height, thereby keeping the film focus distance constant. The film focus
distance shown in the display is the actual distance to the detector. When
Film Tracking
mode
is chosen all buttons except X and alpha-brake buttons will be deactivated. The tube
stand is operated manually by releasing one or both of the brakes. The position of the detec-
tor is changed according to the change in X and or alpha position of the tube. That is the X
and alpha positions can be changed independently.
Exposure Validation
Exposure is allowed if the stand is not moving, operating properly (not in an error state), the
X-ray tube is aimed to the center of the detector and the servo button is activated.
Содержание Omnera 500A
Страница 1: ...Technical Publication Operator s Manual 1000 095 071 EN Rev 1 1 ...
Страница 3: ...Revision history Chapter Pages Rev Date New Release 1 1 2020 09 ...
Страница 4: ......
Страница 62: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 54 2020 09 14 Safety Electromagnetic compatibility EMC ...
Страница 120: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 112 2020 09 14 Fig 4 16 Operating the System Perform examination ...
Страница 152: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 144 2020 09 14 Operating the System Portable detector table ...
Страница 158: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 150 2020 09 14 Error Handling Diagnostic ...
Страница 162: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 154 2020 09 14 Cleaning Wall stand Table OTC and X ray Tube ...
Страница 164: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 156 2020 09 14 Function and safety checks Monthly Checks ...
Страница 172: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 164 2020 09 14 Technical specification Wall stand ...
Страница 174: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 166 2020 09 14 Waste Disposal ...
Страница 186: ......