2.10.1 Radiation protection
Because of the ionizing nature of x-ray radiation, precautions have to be taken to minimize
the harmful effects to patients and operators/staff during exposures. The aim is to achieve
dose levels “as low as reasonable achievable”. National regulatory dose limitation require-
ments have to be followed.
Following four main factors control the amount (dose) of radiation received from a source:
Patient and operator dose:
Loading factors
: Reducing the loading factors reduces the effective dose proportionally.
Lower values will give more noise in the image.
Distance
: Increasing the distance reduces dose levels according to the inverse square law.
Beam size
: Keep the beam size as small as possible.
Shielding
: Whenever possible/necessary protective shielding should be used to limit dose
levels.
2.10.1.1 Protection against primary radiation (patient)
Following measures should to be taken to limit patient dose.
• Observe national dose limit regulations.
• Exposure parameters (time/mA) should be set as low as possible with an acceptable im-
age noise level.
• Set focus to skin distance as large as possible.
• Always collimate the exposure field to the area of interest. This will both decrease the dose
level and improve the image quality (less scattered radiation).
• If possible/necessary use protective shielding.
2.10.1.2 Protection against secondary radiation
As the patient is the most significant source of scattered radiation during an x-ray exam, the
staff and/or operator will unavoidable be exposed to ionizing radiation when inside the x-ray
room during an exposure. Radiation doses from scattered radiation can be significantly high.
Following safety measures should be taken to minimize scattered radiation to the staff.
• Increasing the distance to the central beam reduces dose levels according to the inverse
square law.
• Protective clothing, e.g. lead apron, should always be used.
• Exposure parameters (time/mA) should be set as low as possible.
• Using high kV and low mA produces less scatter.
• Always collimate the exposure field to the area of interest.
• Added collimator filter reduces the scatter.
• Compression of patient.
Содержание Omnera 500A
Страница 1: ...Technical Publication Operator s Manual 1000 095 071 EN Rev 1 1 ...
Страница 3: ...Revision history Chapter Pages Rev Date New Release 1 1 2020 09 ...
Страница 4: ......
Страница 62: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 54 2020 09 14 Safety Electromagnetic compatibility EMC ...
Страница 120: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 112 2020 09 14 Fig 4 16 Operating the System Perform examination ...
Страница 152: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 144 2020 09 14 Operating the System Portable detector table ...
Страница 158: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 150 2020 09 14 Error Handling Diagnostic ...
Страница 162: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 154 2020 09 14 Cleaning Wall stand Table OTC and X ray Tube ...
Страница 164: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 156 2020 09 14 Function and safety checks Monthly Checks ...
Страница 172: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 164 2020 09 14 Technical specification Wall stand ...
Страница 174: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 166 2020 09 14 Waste Disposal ...
Страница 186: ......