![Canon Omnera 500A Скачать руководство пользователя страница 66](http://html.mh-extra.com/html/canon/omnera-500a/omnera-500a_operators-manual_61481066.webp)
OMNERA®
1000-095-071-EN Rev. 1.1
Operator´s Manual
Page 58
2020-09-14
3.2.1 Automatic collimator
60
˚
60
˚
cm
110
ms
20
mA
250
kV
50
Density
1
+
8
mAs
AEC
Jane Doe
DoB 1977-03-06
ID 987-65-4320
Abdomen suspine
Age 36 Sex F
Acc No 987-65-4320
1.0
Al
0.1
Cu
43.0
cm
43.0
cm
x
i
3
2
1
1. Select
Automatic
or
Manual
mode of the collimator.
Note!
If there is no new examination and the System is in a Manual mode, the collimator
ignores the instructions of sent collimator area from the Imaging system. This can be use-
ful if the collimator area is intended to be kept for the next examination.
2. Collimator light/x-ray field height x width
3. Collimator filter selection
3.2.1.1 Collimator mode
4
074
4
074
1
2
Fig. 3-3 Collimator mode
In automatic mode, the collimator light
height and width is set from the anatomical
program. The collimator light size can still
be changed manually. In manual mode the
collimator light height and width is not set
from the anatomical program.
1. Automatic mode
2. Manual mode
When Automatic mode is selected the predefined values of the collimator light / x-ray field
size and the filter selection will be set automatically when the anatomical protocol is selected.
Both the light / x-ray field size and the filter can be changed when needed. In automatic mode
the maximum light / x-ray field size is restricted to the maximum active area of the imaging
unit.
When manual mode is selected the size of the collimator light / x-ray field size can be ad-
justed outside of the maximum active area of the imaging unit. When a new anatomical pro-
tocol is selected (for the same patient) the collimator light / x-ray field size or the filtration is
not changed even if size and filtration is defined different in the protocol.
Содержание Omnera 500A
Страница 1: ...Technical Publication Operator s Manual 1000 095 071 EN Rev 1 1 ...
Страница 3: ...Revision history Chapter Pages Rev Date New Release 1 1 2020 09 ...
Страница 4: ......
Страница 62: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 54 2020 09 14 Safety Electromagnetic compatibility EMC ...
Страница 120: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 112 2020 09 14 Fig 4 16 Operating the System Perform examination ...
Страница 152: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 144 2020 09 14 Operating the System Portable detector table ...
Страница 158: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 150 2020 09 14 Error Handling Diagnostic ...
Страница 162: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 154 2020 09 14 Cleaning Wall stand Table OTC and X ray Tube ...
Страница 164: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 156 2020 09 14 Function and safety checks Monthly Checks ...
Страница 172: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 164 2020 09 14 Technical specification Wall stand ...
Страница 174: ...OMNERA 1000 095 071 EN Rev 1 1 Operator s Manual Page 166 2020 09 14 Waste Disposal ...
Страница 186: ......