2020-09-18
2.10.5 Mechanical Safety, Wall Stand
Note!
The patient or operator is allowed to lean against the Wallstand patient handle or armrest,
but
not
to put any weight on it.
2.10.5.1 General
It is the operator's duty to ensure that any danger to the patient and/or third person is pre-
vented, before movements are released.
2.10.5.2 Standard Version
Possible squeeze hazard areas are indicated in figure. Getting stuck in the imaging unit slide
opening (1) causes squeezing hazard if the imaging unit is moving downward (Z-direction).
Z
1
Max 25kg/55 lbs
Fig. 2-18 Possible squeeze hazards
1. Slide opening of the imaging unit
2.10.5.3 Motorized Wall stand
CAUTION!
It is not allowed to have the patient sitting or standing in the wall stand surroundings,
when the motorized movement is operated.
Содержание OMNERA 400T
Страница 1: ...Technical Publication Operator s Manual 0180 095 418 EN Rev 6 1 ...
Страница 3: ...Rev Date New release 6 1 2020 09 ...
Страница 4: ......
Страница 8: ...Contents OMNERA 0180 095 418 EN Rev 6 1 Operator s Manual Page iv 2020 09 18 ...
Страница 18: ...OMNERA 0180 095 418 EN Rev 6 1 Operator s Manual Page 10 2020 09 18 Introduction System description ...
Страница 50: ...OMNERA 0180 095 418 EN Rev 6 1 Operator s Manual Page 42 2020 09 18 Safety Electromagnetic Compatibility EMC ...
Страница 86: ...OMNERA 0180 095 418 EN Rev 6 1 Operator s Manual Page 78 2020 09 18 User Interfaces Table control elements ...
Страница 148: ...OMNERA 0180 095 418 EN Rev 6 1 Operator s Manual Page 140 2020 09 18 Waste disposal ...
Страница 165: ...Operator s Manual 0180 095 418 EN Rev 6 1 OMNERA 2020 09 18 Page 157 Appendix B Annual Checks ...
Страница 166: ......