5-24
COPYRIGHT © CANON ELECTRONICS INC. 2012
CANON P-208 FIRST EDITION
CHAPTER 5 TROUBLESHOOTING
1. Firm Registration
This mode saves scanner firmware on a
computer for servicing in advance, making it
possible to correctly perform actual firmware
updates.
Operation Procedure
1) On the initial screen, select [Firm Regis-
tration]. Or, select [Registration of Firm]
form the scanner in the task bar of the
initial screen or other screens.
Figure 5-245
2) The following screen is displayed only if
there are no firmware already registered.
Select [OK].
Figure 5-246
3) The firmware registration screen is dis-
played. Select [Register].
Figure 5-247
4) When the file selection screen is dis-
played, select the file.
Note:
The file format is “mot”. The file name
can be changed to anything.