5. Photography
-33-
13
Review the image.
The fundus images and study information are displayed in the control software. Check the images.
To take more images, repeat steps 2 and following.
14
End the study.
End the study with the control software.
15
Lock the stage unit.
Press the stage unit lock button to lock the stage unit.
16
Turn OFF the power of the main unit.
Press the power switch to the
O
side.
Concerning the adjustment of the flash intensity
The flash intensity is already set to the standard level for each photography mode.
To adjust the intensity, press the flash intensity switches.
Each time one of these switches is pressed, the intensity is increased or reduced by steps of 0.3.
The flash intensity selected is shown by the lighted area of the flash intensity indicator.
Concerning locking the stage unit
When the stage unit lock button is pressed, the stage unit can be locked or unlocked in the
following two places:
Standard
Increased by 87% (equivalent to a step of 0.9)
of the standard
Increased by 52% (equivalent to a step of 0.6)
of the standard
Increased by 23% (equivalent to a step of 0.3)
of the standard
Reduced by 19% (equivalent to a step of 0.3)
of the standard
Reduced by 34% (equivalent to a step of 0.6)
of the standard
Reduced by 47% (equivalent to a step of 0.9)
of the standard
Reduced by 56% (equivalent to a step of 1.2)
of the standard
Reduced by 65% (equivalent to a step of 1.5)
of the standard
Lock position
Description
Locked at center
Center position of stage unit. (The unit is locked here upon completion of
the examination.)
Lock at front right
Position where the stage unit has been drawn to the front right.
(This protects the patient’s eyes from unintended operations either while
an examination is being undertaken or after an examination has been
completed.)