CANNON Instrument Company® | Getting Started
15
2.
Carefully insert a thermometer through the access hole in the display hood and into the bath until
the tip of the thermometer is in the middle of the capillary portion of the viscometer tube. The
thermometer passes through the same hole in the top of the bath unit that was used to fill the unit
with fluid. See Figure 9.
Figure 9 — Thermometer Access Holes
3.
Wait for the thermometer to adjust to the temperature of the bath fluid. Read the reference
temperature.
4.
From the main screen, click the current bath temperature button
for the appropriate bath.
The temperature calibration screen opens.
5.
Click the
Reference Temperature
field. See Figure 10.
Figure 10 — Temperature Calibration Screen
6.
A field entry screens opens; enter the reference temperature measured in step 3, then click the
green checkmark to submit it and return to the temperature calibration screen.
7.
Click
Calibrate
, wait for the system to save the adjustment, and then click
Back
to return to the
main screen.
8.
Remove and clean the thermometer.
9.
Repeat steps 2 through 8 for the second bath.
Thermometer
access holes