Fill Baths
Each temperature bath must be filled with Dow Corning 200 silicone fluid prior to operation. The
viscosity of the fluid depends upon the temperature of the bath during most operations. See Table 9 for
bath fluid selection. See Figure 4 for bath location overview.
Table 9 — Silicone Bath Fluid Selection
Bath Operating Range
Recommended Bath Fluid Viscosity
20 °C to 100 °C
10 cSt
101 °C to 150 °C
20 cSt
Figure 4 — Temperature Baths
Warning:
Follow all lab policies regarding the safe handling and disposal
of all hazardous materials.
1.
Open the plumbing lid.
Left bath
Right bath
Display hood
Plumbing lid