New GentleYAG
Cryogen Spray Alignment Procedure
Candela Corporation
8503-01-0845, Revision A
Candela Corporation Proprietary
Page 9 of 10
Results
Comments
Action
If you get
“PURGE
REQUIRED”
This particular
handpiece
passed the
test
You may continue use
of this handpiece
If “PURGE
REQUIRED”
does not
occur
This particular
handpiece
failed the test
Discontinue use of
this handpiece and
call Candela Tech
Support
at (508) –358-
7637 ex 336 or (800) –
733-8550 ex 336
9. Reconnect the Cryogen Connector (see Figure 3) and press reset to use the
system, or repeat the test on other GentleYAG handpiece delivery systems.
10. Depress “purge” may be required until the line is refilled.