Cryogen Spray Alignment Procedure
Candela Corporation
8503-01-0845, Revision A
Candela Corporation Proprietary
Page 5 of 10
Results:
Acceptable alignment-
no further action needed
.
Unacceptable alignment,
repeat the test with a different distance gauge
If the test shows acceptable alignment with the new distance gauge discard the
original distance gauge
If the result is still “unacceptable”, Realign the cryogen nozzle.