New GentleYAG

 Cryogen Spray Alignment Procedure  

Candela Corporation  

 

8503-01-0845, Revision A 

Candela Corporation Proprietary 

 

Page 4 of 10 

Test #1 Cryogen Alignment  

Purpose

:  

 To verify the cryogen spray nozzle is properly aligned with the distance gauge ring  

Procedure:   

1. 

Put the laser in “standby”. Caution- laser should remain in “ Standby” mode 
for duration of test 

2. 

Select “DCD –on”. 

3. 

Install GentleYAG 12mm distance gauge PN 7122-00-3109 

4. 

Select 30ms of DCD Spray.  

5. 

Point hand piece away from objects and personnel (toward the floor). View 
the contact ring of distance gauge, looking from the hand piece. 

6. 

Press purge button.  

-DCD Spray should flow completely through the contact ring.  There may be a 
minimal spray mist seen hitting the contact ring. No spray should be spraying 
beyond outside of the contact ring.

 

Acceptable-No mist on contact ring

 

 

 

Acceptable-Small amount of mist on contact 

 

 

Unacceptable results-Excess mist on contact ring 

   

 

 
 
 

No Cryo clipping on 12 
mm distance gauge 

Cryo clipping on 12 
mm distance gauge 

Small Cryo clipping on 12 
mm distance gauge

 

Содержание New GentleYAG

Страница 1: ...New GentleYAG Cryogen Spray Alignment Procedure Candela Corporation 8503 01 0845 Revision A Candela Corporation Proprietary Page 1 of 10 New GentleYAG Cryogen Spray Alignment Procedure...

Страница 2: ...late 1301 00 8291 available from Candela Corporation VPYAG distance gauge 050 inch Allen Wrench The following tests are described in this section 1 Cryogen Alignment Verifies the cryogen spray nozzle...

Страница 3: ...New GentleYAG Cryogen Spray Alignment Procedure Candela Corporation 8503 01 0845 Revision A Candela Corporation Proprietary Page 3 of 10...

Страница 4: ...3109 4 Select 30ms of DCD Spray 5 Point hand piece away from objects and personnel toward the floor View the contact ring of distance gauge looking from the hand piece 6 Press purge button DCD Spray s...

Страница 5: ...oprietary Page 5 of 10 Results Acceptable alignment no further action needed Unacceptable alignment repeat the test with a different distance gauge If the test shows acceptable alignment with the new...

Страница 6: ...ended to represent treatment parameters but rather provide a check on proper functionality of the handpiece and provide a reference for the user to help identify changes in the handpiece operation Pro...

Страница 7: ...Proprietary Page 7 of 10 Distance gauge spot size Typical spray duration to minimally fill the inside ring spot 3 mm 30 ms 8 mm 30 ms 10 mm 30 ms 12 mm 40 ms 15 mm 50 ms 18 mm 60 ms Results If the spo...

Страница 8: ...empty Warning Put on appropriate laser safety glasses 2 Turn on the system and allow the WARM UP to complete 3 With the Electrical Connector installed disconnect the Cryogen Connector see Figure 3 El...

Страница 9: ...You may continue use of this handpiece If PURGE REQUIRED does not occur This particular handpiece failed the test Discontinue use of this handpiece and call Candela Tech Support at 508 358 7637 ex 336...

Страница 10: ...New GentleYAG Cryogen Spray Alignment Procedure Candela Corporation 8503 01 0845 Revision A Candela Corporation Proprietary Page 10 of 10...

Отзывы: