Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
operating room should install air-conditioning equipment to ensure the
temperature about 25
℃
before using the machine; otherwise the water
temperature will be a sharp rise in a short time, which will not all shortens the
working time of the machine , but also affect the service life of the machine.
d. Water quality monitoring. Machine Built-in cooling water quality monitoring
module. if the water is impurities, and water quality is bad, the machine will
no longer monitor the water flow, suggesting that "water is not circulating." It
should replace the cooling water in this time.
e. water level monitoring. The machine has no built-level monitoring module or
device.
External water level displays the current flow in the water tank. The water line
should be 7 or more.