Appendix C. Calibration Wizard Tutorial
C-2
4.
A waveform with a default window length of 50 m is then displayed with
PC-TDR
highlighting where it thinks the probe is found. If the probe is not
within the highlighted region, click on the highlighted area in the wizard’s
window and move or scale the region to encompass the entire portion of
the probe. If it is difficult to find the probe or the probe was not entirely in
water prior to the measurement, select
Refresh
button and measure the
probe again. Once the cable and window lengths are satisfactory, click
Next
.
5.
Enter the temperature of the water and select
Next
.