![Campbell SR50AT Скачать руководство пользователя страница 75](http://html1.mh-extra.com/html/campbell/sr50at/sr50at_instruction-manual_488987075.webp)
Appendix E.
Tera Term
QuickStart
E-7
1. Ensure the SR50AT/T is completely disconnected from power or
communication. The best way to do this is to simply unscrew the cable
connector from the SR50AT/T.
2. Open the SR50AT/T as follows:
a. Remove the six Phillips screws (slot-headed screws in older models)
that are located on the bottom side of the transducer housing.
b. The transducer housing will separate from the main metal housing.
c. Ensure that the screws are kept and set aside. On older models, the
O-ring is removable. Remove the O-ring and set it aside as well.
d. Disconnect the transducer housing assembly from the main metal
body by pressing down on the tab on the connector and gently pulling
the connectors apart. You may have to gently rock the connector from
side to side to get it to separate.
e. Remove the two Phillips screws from the bottom of the disk assembly.
Notice that these are the only two countersunk holes in the disk
assembly. Set the two screws aside.
f. The bottom disk assembly and the transducer housing have a
V-shaped notch that lines up with a mating V in the metal housing.
Separate the bottom disk from the main metal housing.
g. The circuit board will remain connected to the housing via the signal
wires.
h. Place the bottom disk/circuit assembly on the work bench with the
circuit board and DB-9 connector facing up. Ensure that no part of the
PCB is in contact with the metal lid or other conductive objects.
3. Move the jumper on the RUN/PROG header from the RUN position to the
PROG (program) position.
4. Connect to the DB-9 connector on the SR50AT/T PC board.
5. Connect the cable back to the SR50AT/T. The cable should be providing
the SR50AT/T with power.
6. When power is reestablished, you will see a green LED light up close by
the RUN/PROG jumper. The LED will be on continuously with no
flashing.
7. Connect to the SR50AT/T with
Tera Term
.
The SR50AT/T will not accept any commands when in the
program mode.
8. In
Tera Term
select
File
at the top of the screen.
9. From the drop-down menu select
Send file
…
.
NOTE