SR50AT-Series Sonic Ranging Sensors
5
6. Specifications
Features:
•
Wide operating temperature range
•
Compatible with Campbell Scientific CRBasic dataloggers:
CR200(X) series, CR300 series, CR6 series, CR800 series, CR1000,
CR3000, and CR5000
Power Requirements:
9 to 18 Vdc
Quiescent Power
Consumption (no heater)
SDI-12 Mode:
< 1.0 mA
RS-232/RS-485 Modes:
< 1.25 mA (≤ 9600 bps),
< 2.0 mA (> 9600 bps)
Active Power
Consumption (no heater):
250 mA typical
Measurement Time:
Less than 1.0 s typical for RS-232 or
RS-485 measurements.
1.2 s typical for SDI-12 measurements.
Selectable Outputs:
SDI-12 (version 1.3)
RS-232 (1200 to 38400 bps)
RS-485 (1200 to 38400 bps)
Measurement Range:
0.5 to 10 m (1.6 to 32.8 ft)
Accuracy:
±1 cm (±0.4 in) or 0.4% of distance to
target, whichever is greater.
Accuracy specification excludes errors in
the temperature compensation. An external
temperature compensation is required for
the SR50AT.
Resolution:
0.25 mm (0.01 in)
Required Beam Angle Clearance:
30°
Operating Temperature Range:
–45 to 50 °C
Maximum Cable Length:
SDI-12 60 m (196.9 ft)
RS-232 (9600 bps or less) 30 m (98.4 ft)
RS-485 300 m (984.3 ft)
1
Cable Type:
4 conductor, 2-twisted pair, 22 awg,
Santoprene jacket
Temperature Sensor
Cable Length:
2.74 m (9 ft)