![Campbell SkyVUE 8 Скачать руководство пользователя страница 45](http://html1.mh-extra.com/html/campbell/skyvue-8/skyvue-8_product-manual_3808856045.webp)
Table 5-1: Summary of the terminal mode commands available
Command
Parameter/
parameter block
Description
full settings from one SkyVUE 8 to another with the
SETUSER
command.
A typical response to the
GETUSER
command is:
>>>>> COPY FROM START OF NEXT LINE >>>>>
user 7 007638-6da 106 510 0 , 0 10 2 0 1 14520 400 1
1000 1000 2000 0 1 0 0 0 0 10 0.000E+00 2.000E+00 1
1.000E+00 0 1.000E+00 0 10 1 0 10 0 100 4.000E+00
6.000E+00 2.000E-03 2.500E-04 1.000E+03 2.200E+01
4.800E+03 1.000E-01 9.000E-01 2.500E-04 2.000E+03
4.500E+01 2.000E+00 50 2.00E-03 3.000E+01
1.500E+02 1.800E+00 5.000E+00 1.000E+01 150 98b2
<<<<< TO START OF THIS LINE <<<<<
HEATERS
Hood
Internal
Laser
Test_interval
Sets or reads heater settings as follows:
Hood
= 0, Hood blower and heater OFF
Hood
= 1, Hood blower ON and heater OFF
Hood
= 2, Hood blower ON and heater ON
Hood
= 3, Hood blower and heater AUTO (default) –
(see note (1) below).
Internal
= 0, Internal heater OFF
Internal
= 1, Internal heater ON
Internal
= 2, Internal heater AUTO (default)
Laser
= 0, laser heater off
Laser
= 1, laser heater on (default)
Test_interval
= 1 to 168 hours (default 24
hours). Heater/Blower test interval
HELP
No parameters
Calls up a list of user commands with brief
descriptions
HOFFSET
Height_offset
Height_offset
is the offset to be added or
subtracted in the range of ±1000 m (±3281 ft). Positive
SkyVUE™8 (CS136) LIDAR Ceilometer