Table 5-1: Summary of the terminal mode commands available
Command
Parameter/
parameter block
Description
ALARMS
Angle
Angle is the threshold tilt angle of the sensor beyond
which an alarm will be flagged. The settable range is 0
to 90.0 degrees tilt and the default is 45.0 degrees.
APPLICATION
Application_n
Application_n
selects pre-configured user
settings optimized for a specific application. For the
following values of n:
0 = Help
10 = Aviation
20 = Research
40 = Meteorology
225 = User defaults
The default application for the SkyVUE 8 is 10,
Aviation. Settings adjusted by the
APPLICATION
command include settings in
BS
,
MCFG
, and
UNITS
.
BS
(see
Measurement
and message
intervals
for more
information on
compatibility of
different
parameters)
Attenuated_
SCALE
,
BS_Av_Time
,
Noise_Gate
,
Measurement_
Period
,
Rolling_
Average
,
Message_
Interval
BS_Temporal_
Filter_Mode
Rules for
BS
command are:
BS_Av_Time <= Measurement_Period-1s
.
Message_Interval
must be a multiple of
Measurement_Period
.
Attenuated_SCALE
is the scalar for the attenuated
backscatter percent. 0.001 to 1000% (default 100%)
BS_Av_Time
is the backscatter average time in
seconds. 1 to 30 (default 7).
Note: Must not be greater than
Measurement_
Period
– 1.0
Noise_Gate
controls the noise threshold applied to
back-scatter.
Noise_Gate
= -1000, all backscatter range is
corrected.
Noise_Gate
= 0.0 to 1000.0, standard deviation
SkyVUE™8 (CS136) LIDAR Ceilometer
28