NR01 Four-Component Net Radiation Sensor
9
6.3.1 Using a 4WPB100
The CR1000 and CR9000(X) require the 4WPB100 module (FIGURE
) to
interface the PT100 to the datalogger. Cable 1 is used to connect the
pyranometer and pyrgeometer (TABLE
and TABLE
). Cable 2
connects the PT100 to the 4WPB100 and datalogger (TABLE
).
FIGURE 6-2. 4WPB100 Module
If free channels are limited, it is possible to measure the PT100
sensor using a 3WHB10K terminal input module, with only a
slight loss of accuracy. This only requires one differential
channel. Please refer to the documentation for that module for
further details.
TABLE 6-2. Cable 1-to-Datalogger Connections when using
Differential Measurements and the 4WPB100*
Wire Label
Color
CR1000, CR9000(X)
Pyranometer Up Sig
Red
Differential Input (H)
Pyranometer Up Ref
**Blue
Differential Input (L)
Pyranometer Down Sig
White
Differential Input (H)
Pyranometer Down Ref
**Green
Differential Input (L)
Pyrgeometer Up Sig
Brown
Differential Input (H)
Pyrgeometer Up Ref
**Yellow
Differential Input (L)
Pyrgeometer Down Sig
Purple
Differential Input (H)
Pyrgeometer Down Ref
**Grey
Differential Input (L)
Shield
Shield
⏚
*Ensure that it is cable 1 before connecting.
**Jumper to
⏚
with user supplied wire.
NOTE