NR01 Four-Component Net Radiation Sensor 

TABLE 6-1.  Recommendations for Installation of the NR01 

Location  

Location of measurement should be representative of the 

total surrounding area, in particular in case the NR01 is used 

for environmental net radiation measurements.  If possible, 

mount the sensor on a separate pole at least 25 ft away from 

main logger tower or tripod. 

Mechanical 

mounting  

A 2-axis leveling assembly is included as part of the sensor 

mount which is suitable for a range of pipe diameters, max 

1-inch IPS pipe (33 mm diameter).  

Radiation 

detection  

Avoid objects that cast shadows or reflections on the 

instrument.  This includes both incoming and outgoing 

radiation. 

Leveling  

Use the bubble-level to see if the instrument is mounted 

horizontally.  For viewing the level, the radiation screens 

must be removed.  Alternatively a spirit level can carefully 

be put on the pyrgeometer window.  

Orientation  

By convention, with the wiring to the nearest pole (north in 

the northern hemisphere, south in the southern hemisphere). 

Height of 

installation  

In case of inverted installation, a height of approximately 4 

ft (1.5 m) above ground is recommended by the WMO (to 

get good spatial averaging). 

Tilt  

The NR01 should typically be installed horizontally, but for 

some applications, may be installed in a tilted position.  In 

all cases, it will measure the fluxes that are incident on the 

surface that is parallel to the sensor surface.  

 

6.3  Datalogger Wiring 

The four radiation outputs can be measured using differential or single-ended 

inputs on the datalogger.  A differential voltage measurement is recommended 

because it has better noise rejection than a single-ended measurement.  When 

differential inputs are used, jumper the low side of the input to 

 to keep the 

signal within the common mode range. 

Cables generally act as a source of signal distortion by picking up capacitively 

coupled noise.  Therefore, Campbell Scientific recommends keeping the 

distance between the datalogger and sensor as short as possible.  

The sensor has two cables with similar color schemes.  

It is important to make sure you identify cable 1 and 

cable 2 correctly, especially before connecting any 

source of power such as to the heater.  Failure to do so 

may damage the sensor. 

WARNING 

Содержание NR01

Страница 1: ...INSTRUCTION MANUAL NR01 Four Component Net Radiation Sensor Revision 9 15 C o p y r i g h t 2 0 0 8 2 0 1 5 C a m p b e l l S c i e n t i f i c I n c...

Страница 2: ......

Страница 3: ...ucts to CSI CSI will return such Products by surface carrier prepaid within the continental United States of America To all other locations CSI will return such Products best way CIP port of entry per...

Страница 4: ...g container Campbell Scientific s shipping address is CAMPBELL SCIENTIFIC INC RMA _____ 815 West 1800 North Logan Utah 84321 1784 For all returns the customer must fill out a Statement of Product Clea...

Страница 5: ...beginning work Wear a hardhat and eye protection and take other appropriate safety precautions while working on or around tripods and towers Do not climb tripods or towers at any time and prohibit cl...

Страница 6: ......

Страница 7: ...er 11 6 5 Installation of the Radiation Shields 12 6 6 Datalogger Programming 12 6 6 1 Pyranometer Pyrgeometer Measurements 12 6 6 2 Calibration Factors 13 6 7 Instrument Inversion Test 13 7 Operation...

Страница 8: ...Specifications 6 6 1 Recommendations for Installation of the NR01 8 6 2 Cable 1 to Datalogger Connections when using Differential Measurements and the 4WPB100 9 6 3 Cable 1 to Datalogger Connections...

Страница 9: ...ons READ AND UNDERSTAND the Safety section at the front of this manual Care should be taken when opening the shipping package to not damage or cut the cable jacket If damage to the cable is suspected...

Страница 10: ...in the pyrgeometers connection body and is used to prevent dew Measurement of the separate components of the net radiation is useful because it Enhances accuracy by having separately calibrated instr...

Страница 11: ...emperature and off course net radiation net value of all SW and LW fluxes can be calculated The NR01 requires leveling a two axis leveling facility is incorporated in the design See Section 6 Installa...

Страница 12: ...for 12 hour totals day and night Operating temperature 40 to 80 C Sensitivity All sensors have individual calibration factors Tpyrgeometer PT100 DIN class A Tpyrgeometer accuracy Within 1 C Tpyrgeome...

Страница 13: ...Non Linearity 2 5 Directional response for beam radiation Within 25 W m2 Spectral selectivity 5 305 to 2000 nm Temperature response within an interval of 50 C Within 6 10 to 40 C Tilt response Within...

Страница 14: ...f view 150 degrees Spectral selectivity Not specified Temperature response within an interval of 50 C Within 6 10 to 40 C Tilt response Within 2 Sensitivity 5 15 V Wm 2 Expected voltage output Meteoro...

Страница 15: ...or pyranometer A level is located under the radiation screens 6 2 Mounting The NR01 has a hole suitable for a 1 inch IPS pipe 33 mm diameter for mounting the NR01 onto a CM204 or CM206 crossarm The cr...

Страница 16: ...ere Height of installation In case of inverted installation a height of approximately 4 ft 1 5 m above ground is recommended by the WMO to get good spatial averaging Tilt The NR01 should typically be...

Страница 17: ...quires one differential channel Please refer to the documentation for that module for further details TABLE 6 2 Cable 1 to Datalogger Connections when using Differential Measurements and the 4WPB100 W...

Страница 18: ...rown Single Ended Input Pyrgeometer Up Ref Yellow Pyrgeometer Down Sig Purple Single Ended Input Pyrgeometer Down Ref Grey Shield Shield Ensure that it is cable 1 before connecting TABLE 6 4 Cable 2 C...

Страница 19: ...put H PT100 Signal Ref Green U configured as a differential input Differential Input L Current Excite Red U configured as a current excitation Current Excitation IX Current Return Blue U configured as...

Страница 20: ...asic dataloggers are provided in the following sections Equations used to calculate solar parameters and measurement details are provided in Section 7 Operation p 13 Complete program examples for a CR...

Страница 21: ...nverting the instrument position 180 degrees turn and looking at the output signals The instrument output should have the same magnitude but a reversed sign so to and to For best results perform this...

Страница 22: ...radiation hits the sensor perpendicularly normal to the surface sun at zenith 0 degrees angle of incidence zero response occurs when the sun is at the horizon 90 degrees angle of incidence 90 degrees...

Страница 23: ...sky For down facing instruments the earth surface reflects part of the solar radiation depending on the local surface properties If there is direct radiation this often is the dominant source of ener...

Страница 24: ...lled directional response or cosine response that is close to the ideal cosine characteristic To attain the proper directional and spectral characteristics a pyrgeometer s main components are 1 Thermo...

Страница 25: ...components 1 Low temperature radiation from the universe filtered by the atmosphere The atmosphere is transparent for this radiation in the so called atmospheric window around 10 to 15 micrometer wav...

Страница 26: ...radiation values in TABLE 7 3 are corrected for sensor temperature The values in TABLE 7 4 are not corrected for sensor temperature On a smaller timescale the most important factors are solar position...

Страница 27: ...c recommends activating the heater to avoid dew deposition During the day the window heating offset may affect the downward facing pyrgeometer s output This can produce a maximum error of 15 W m2 At n...

Страница 28: ...nternal connections see TABLE B 3 The sensor signal is unrealistically high or low Check if the right calibration factors are entered into the algorithm Please note that each sensor has its own indivi...

Страница 29: ...vg SR01 Down short wave radiation Avg IR01 Up long wave radiation Avg IR01 Down long wave radiation Avg NR01 temperature degrees C Avg NR01 temperature degrees K Avg Net shortwave radiation Avg Net lo...

Страница 30: ...Public IR01UpCo Public IR01DnCo Units Batt_Volt Volts Units SR01Up W m2 Units SR01Dn W m2 Units IR01Up W m2 Units IR01Dn W m2 Units NR01TC Deg C Units NR01TK K Units NetRs W m2 Units NetRl W m2 Units...

Страница 31: ...lbedo UpTot DnTot NetTot IR01UpCo and IR01DnCo For the CR1000 use autorange for the SR01 measurements due to the wide dynamic range VoltDiff SR01Up 1 autorange 1 True 0 _50Hz SR01UpCal 0 VoltDiff SR01...

Страница 32: ...o Avg Total Net radiation Avg temperature corrected IR01 Up Avg temperature corrected IR01 Down CR3000 Datalogger ANALOG INPUT 1H SR01 UP downwelling shortwave radiation signal red 1L SR01 UP downwell...

Страница 33: ...False Average 1 NetRl FP2 False Average 1 Albedo FP2 False Average 1 UpTot FP2 False Average 1 DnTot FP2 False Average 1 NetTot FP2 False Average 1 IR01UpCo FP2 False Average 1 IR01DnCo FP2 False EndT...

Страница 34: ...temperature RH probe is not used The algorithm turns the heater on off at 4 Hz The duty cycle of the pulse is changed depending on how close the radiometer body temperature is to the dewpoint tempera...

Страница 35: ...5 Const NR01_SHORT_UW_CAL 1000 13 93 Unique NR 01 shortwave upwelling multiplier 1000 13 5 Const NR01_LONG_DW_CAL 1000 8 8 Unique NR 01 longwave downwelling multiplier 1000 10 5 Const NR01_LONG_UW_CAL...

Страница 36: ...ample 1 rh_hmp_mean IEEE4 Sample 1 t_dew_hmp_mean IEEE4 Sample 1 duty_cycle IEEE4 Average 9 Rn IEEE4 FALSE EndTable DataTable scratch TRUE 1 TableHide DataInterval 0 5 Min 10 Average 1 t_hmp IEEE4 FAL...

Страница 37: ...using 4 Hz pulse width modulation Below the dew point temperature applies 100 power Above the dew point power is reduced linearly to 20 until the dew point plus DELTA_SET_POINT_1 After the dew point...

Страница 38: ...Appendix A Example Programs A 10...

Страница 39: ...Pyranometer Down Sig 8 Green Pyranometer Down Ref 7 Brown Pyrgeometer Up Sig 4 Yellow Pyrgeometer Up Ref 3 Purple Pyrgeometer Down Sig 6 Grey Pyrgeometer Down Ref 5 Clear Shield Ground 11 12 The conne...

Страница 40: ...f the NR01 for servicing purposes only PCB04 Connection PCB04 Terminal PCB05 Connection PCB05 Terminal Polarity 3 8 Pyrgeometer UP 4 7 Pyrgeometer UP 5 12 Pyrgeometer DOWN 6 11 Pyrgeometer DOWN 13 Pyr...

Страница 41: ......

Страница 42: ...ua Apinag s nbr 2018 Perdizes CEP 01258 00 S o Paulo SP BRASIL www campbellsci com br vendas campbellsci com br Campbell Scientific Canada Corp 14532 131 Avenue NW Edmonton AB T5L 4X4 CANADA www campb...

Отзывы: