SECTION 6. MAINTENANCE AND TROUBLESHOOTING
6-3
6.2.3 -99999 DISPLAYED IN AN INPUT
LOCATION
A.
Make sure the battery voltage is between
9.6 and 16 VDC.
B.
With the MetData1 or ET106, verify that the
sensor is connected to the proper bulkhead
connector. With custom weather stations,
verify the sensor is wired to the analog
channel specified in the measurement
instruction or Short Cut .FSL file (single-
ended channels are not labeled on the older
silver-colored wiring panels and are
numbered sequentially starting at 1H; i.e. 1L
is single-ended channel 2).
C. Make sure the Range parameter in the
measurement instruction covers the full
scale voltage output by the sensor.
6.2.4 UNREASONABLE RESULTS DISPLAYED
IN AN INPUT LOCATION
A.
Inspect the sensor for damage and/or
contamination.
B.
Make sure the sensor is properly wired to
the datalogger.
C. Check the multiplier and offset parameters
in the measurement instruction.
6.2.5 6999 OR 99999 STORED IN FINAL
STORAGE (OR STORAGE MODULE)
A. Final Storage format limitations exceeded
(any number larger than 6999 in low
resolution, or 99999 in high resolution
format is stored as the maximum number).
Change the datalogger program.
Содержание ET101