CS135 Ceilometer
46
CR LF
= Carriage Line Feed
LINE 3
scale
9
res
9
n
9
energy
9
lt
9
tr
9
ti
_
bl
9
L0112HN15
9
sum
CR LF
where
scale
(5 characters) = Scale parameter,%, 0 ... 99999. 100% is typical
res
(2 characters) = Backscatter profile resolution in metres.
n
(4 characters) = Profile length 385, 770, 1400, or 1500 samples.
energy
(3 characters) = Laser pulse energy, %.
lt
(3 characters including l/-) = Laser temperature, degrees C
tr
(3 characters) = Window transmission, %
ti
(2 characters) = Total tilt angle, degrees
bl
(4 characters) = Background light, millivolts at internal ADC input (0 ... 2500)
"L0112HN15" (9 characters) = Reserved, defaults shown
sum
= Sum of detected and normalized backscatter, 0 ... 999. Not used in Operating System
1
CR LF
= Carriage Line Feed
This line is omitted from message 105.
LINE 4
SSSSSSSSSSSSSSSSSSSSSSSSSSSSS........(5 x 770 bytes)
CR LF
The two-way attenuated normalised backscatter profile
(100000·srad·km)
-1
unless otherwise scaled by the
Attenuated
_
SCALE
parameter. Each sample is
coded with a 20-bit HEX ASCII character set; msb nibble and bit first, 2's complement.
The profile is not corrected for the tilt angle.
CR LF
= Carriage Line Feed
This line is omitted from message 105.
NOTE
NOTE
NOTE