CS115 Barometric Pressure Sensor

6

Once the CS115 is turned on one minute before every hour, and the correct
measurement is made, it is then copied into the current variable called
"Pressure".  The sensor is then turned off immediately following the
measurement.  In this example, the CR5000 measures the CS115 once every
hour while measuring other sensors at 10 Hz.

In Example 5, the 21X turns on the CS115 one minute before the top of the
hour, and on the hour it measures the CS115 using the pulse count instruction.
The execution interval should be set at 1 second, and the configuration code for
Pulse instruction  (P3) should be set at 24, Low Level AC, 16-bit, Output in Hz.
When this option is selected, two adjacent 8-bit pulse counters (Pulse Input
channels 1 and 2) are combined into one input channel (Pulse Input channel 1)
to form a 16-bit accumulator.  Therefore, Pulse Input channel 2 cannot be used.
You can also choose to use Pulse Input channel 3, in which case Pulse Input
channels 4 will be combined with Pulse Input channel 3 to form a 16-bit
accumulator.  In this mode, the 21X’s execution interval 

must

 be less than one

minute.  If the execution interval is longer than one minute, you will get an
error, because the 16-bit pulse accumulator will overflow counting the signal
from the CS115 when it outputs close to 1100 Hz (1100 mb of barometric
pressure).

In the example programs above, the pressure is reported in millibars (mb).  To
report pressure in different units, multiply the measured pressure by the
appropriate conversion factor using the P37 (Z=X*F) instruction for the
CR500, CR510, CR10(X), CR23X, 21X and CR7, or by adding an expression
for the CR5000 and CR9000.  See Table 5-1 below for conversion factors.

TABLE 5-1.  Conversion Factors for

Alternative Pressure Units

To Find

Multiply by

hPa

1.0

kPa

0.1

mm of Hg

0.75006

in of Hg

0.02953

psi

0.0145

atm

0.00099

torr

0.75006

Содержание CS115

Страница 1: ...INSTRUCTION MANUAL CS115 Barometric Pressure Sensor 3 03 C o p y r i g h t c 2 0 0 2 2 0 0 3 C a m p b e l l S c i e n t i f i c I n c...

Страница 2: ...l other warranties expressed or implied including warranties of merchantability or fitness for a particular purpose CAMPBELL SCIENTIFIC INC is not liable for special indirect incidental or consequenti...

Страница 3: ...Period Averaging Measurement Example for CR10 X 7 5 1 2 Period Averaging Measurement Example for CR10 X in a Slow Executing Program 8 5 1 3 Period Averaging Measurement Example for CR23X 9 5 1 4 Perio...

Страница 4: ...This is a blank page...

Страница 5: ...cting as a pressure sensitive diaphragm and the other acting as a resonator Pressure variations deflect the sensitive diaphragm and change the sensor s resonant frequency The resonant frequency is mea...

Страница 6: ...rnal Trigger Voltages ON 1 to 24 VDC OFF 0 VDC Warm up Time 2 seconds 40 C to 60 C EMC Compatibility Emissions BS EN50081 1 Immunity BS EN61000 6 2 2 3 Physical Dimensions Main Box 2 36 x 2 36 x 1 15...

Страница 7: ...rmanent damage If it is necessary to make a vent hole on the outer wall of an enclosure do not make the hole on one of the vertical side walls as wind blowing around it can cause transient changes in...

Страница 8: ...atalogger s control port In this mode power consumption is minimized but the sensor is subject to a 2 second warm up delay after it is turned on You can also configure the unit for the continuous mode...

Страница 9: ...truction e g 21X CR7 or CR9000 a pulse count instruction can be used to measure the CS115 Atmospheric pressure changes little with time In most weather station applications measuring pressure once an...

Страница 10: ...accumulator Therefore Pulse Input channel 2 cannot be used You can also choose to use Pulse Input channel 3 in which case Pulse Input channels 4 will be combined with Pulse Input channel 3 to form a 1...

Страница 11: ...et Port 8 High Measure CS115 at the top of the hour 2 If time is P92 1 0 Minutes Seconds into a 2 60 Interval same units as above 3 30 Then Do 3 Period Average SE P27 1 1 Reps 2 14 200 kHz Max Freq 2...

Страница 12: ...0 X in a Slow Executing Program EXAMPLE 2 Sample Program for CR10 X CR10X Table 1 Program 01 10 Execution Interval seconds Turn on CS115 1 Do P86 1 48 Set Port 8 High Give 2 second delay for CS115 to...

Страница 13: ...utines End Program Input Locations 1 P_mb Proper entries will vary with program and datalogger channel and input location assignments 5 1 3 Period Averaging Measurement Example for CR23X EXAMPLE 3 Sam...

Страница 14: ...Do P86 1 51 Set Port 1 Low 5 End P95 6 If time is P92 1 0 Minutes Seconds into a 2 60 Interval same units as above 3 10 Set Output Flag High Flag 0 7 Real Time P77 1 0110 Day Hour Minute midnight 000...

Страница 15: ...r If IfTime 59 60 min Then WriteIO 1 1 Period Avgeraging instruction must be executed every scan and cannot be inside the If statement The measured value therefore is first written into a temporary va...

Страница 16: ...If time is P92 1 0 Minutes into a 2 60 Minute Interval 3 30 Then Do 3 Pulse P3 1 1 Reps 2 1 Pulse Input Channel 3 24 Low Level AC 16 Bit Output Hz 4 1 Loc P_mb 5 1 Mult 6 0 0 Offset 4 Do P86 1 56 Set...

Страница 17: ...1 mb resolution you either need to deduct a fixed offset from the reading before it is stored to avoid exceeding the 699 9 or 799 9 threshold or output the barometric reading in high resolution forma...

Страница 18: ...connection and pipe are secure and undamaged The external case can be cleaned with a damp lint free cloth and a mild detergent solution Druck recommends recalibrating the CS115 every year Contact Cam...

Страница 19: ...This is a blank page...

Страница 20: ...campbellsci com br suporte campbellsci com br Campbell Scientific Canada Corp CSC 11564 149th Street NW Edmonton Alberta T5M 1W7 CANADA www campbellsci ca dataloggers campbellsci ca Campbell Scientifi...

Отзывы: