 8.  Click the Live Data tab to view real-time measurements. 

The Quality Metric Sample value should be greater than 
5.

WARNING:
Remove the USB cable before closing the door.

4. Wire functions and connections

Connect the wires in the order shown in the following table.

Table 4-1: Wire color, function, and data logger 
connection

Wire color

Wire function

Data logger connection

White

SDI-12 signal

C, SDI-12, or U configured 

for SDI-12

1

Red

Power

12V, 

Black

Power ground

G

Blue

RS-232 TX

not connected

Yellow

RS-232 RX

not connected

1

U and C terminals are automatically configured by the 

measurement instruction.

5. Mounting

CAUTION:
To prevent condensation, install the sensor in an 
environmentally protected enclosure, complete with 
desiccant. Change desiccant at regular intervals.

The BaroVUE 10 is typically mounted in a Campbell Scientific 
enclosure next to the data logger. Secure the BaroVUE 10 to an 
enclosure backplate by inserting screws through the mounting 
holes in the BaroVUE 10 and into the grommets.

FIGURE 5-1. BaroVUE 10 mounted in a Campbell 

Scientific enclosure