AM25T Solid-State Thermocouple Multiplexer
9
FIGURE 7-2. Differential Measurement of Type T Thermocouple
7.3.2 Differential Analog Measurements
Connect one differential sensor to a differential AM25T input channel.
Connect the sensor shields to the analog ground (
⏚
) terminals next to the input
channel. Up to 25 differential sensors may be measured by one differential
channel on the data logger (FIGURE
FIGURE 7-3. Differential Voltage Measurement
7.3.3 Mixed Sensor Types
Different sensors may be mixed on the AM25T. Additional loops and
measurement instructions will be required.
8. Operation
8.1 General Measurement Considerations
Wires have additional capacitance that increases the time required for a signal
to settle its true value. To reduce settling time, Campbell Scientific
Control/Power
Datalogger
Copper
Copper
GND
1H
1L
VX
AM25T
H
L
VX
Analog
Ground
1
1H
1L
2H
2L
CU
CO
Shield
CU
CO
Shield
T
1
T
2
1
Analog Ground is the
terminal between
VX
and
Hi
on the AM25T
Control/Power
Datalogger
Copper
Copper
GND
1H
1L
G
AM25T
H
L
1H
1L
2H
2L
S
S
Shield
Shield
G
1
Analog Ground is the
⏚
terminal between
VX
and
HI
on the AM25T.