c. Click the Applybutton. ALERT2 administrator is now ready to receive data from the

serial port of the AL205R/AL205B.

7. Start A2Control and connect to your AL205R/AL205B (see

Configuration

(p. 5)).

8. Click the gear icon to access the configuration settings.

9. If using TCP/IP, do the following:

a. Click the IP Settings tab.

b. In the ASCII IP Forward field, type the DNS or IP address of the LoggerNet Admin

server, followed by a colon, then the TCP Callback Port. For example, in the following
screen, 192.168.91.130 is the IP address for the LoggerNet Admin server and 62531 is
the TCP Callback Port.

AL205R / AL205B ALERT2 Repeater / Base

17

Содержание AL205B

Страница 1: ...09 2020 Copyright 2020 Campbell Scientific Inc...

Страница 2: ...Transmitting data 9 3 2 Receiving data 9 3 3 Pass reject list 10 4 Firmware updates 10 4 1 Local update 11 4 2 Remote update 12 5 Downloading ALERT2 data 12 5 1 Remote download 13 5 2 Local download U...

Страница 3: ...RT2 standards maintained by the National Hydrologic Warning Council NHWC and the ALERT2 Protocol Technical Working Group TWG formally the ALERT Users Group AUG The ALERT2 protocol was released by the...

Страница 4: ...r called decoder_output on the USB flash drive If and only if the device finds this folder ALERT2 output will be synchronized to the folder The USB flash drive is also used to install a firmware updat...

Страница 5: ...P server The AL205R AL205B will not transmit using TDMA with an NTP time source GPS On Lights up to indicate the GPS is currently powered up The GPS may require several minutes to initially get a fix...

Страница 6: ...v1 1 API v1 0 Supports ALERT2 API input and ALERT2 API and ASCII output Implements proposed protocol extension for Configurable Forward Error Correction Encryption and Authorization and End to End Re...

Страница 7: ...ol 1 4 0 exe version with an AL205R AL205B using firmware version 1 4 0 Previous and current versions of A2Control are available at www campbellsci com downloads term A2Control b 4 Both the AL205R AL2...

Страница 8: ...n the serial port drop down menu instead of a port number Remote access to the AL205R AL205B is disabled by default To enable it launch A2Control software click the gear icon click the IP Settings tab...

Страница 9: ...for the ALERT2 network The frame needs to be long enough to accommodate all remote stations repeaters and possible future stations transmitting on the same frequency TDMA Slot Length IND Settings Sets...

Страница 10: ...er used in the Airlink preamble The value is in milliseconds RF Tail Time IND Settings The amount of unmodulated carrier to follow after the transmitted frame The value is in milliseconds Invert Modul...

Страница 11: ...decodes ALERT2 data when powered When the AL205R AL205B finds a valid Bit Sync pattern in the incoming audio it turns on the Bit Sync LED and begins listening for a valid Frame Sync pattern which sho...

Страница 12: ...if they are in the list Messages can be filtered by Source Address Destination Address or the path list added by other repeaters Messages can either be reported locally in the ALERT2 logs or dropped s...

Страница 13: ...humb drive into the USB port of the AL205R AL205B 4 Wait more than two minutes for the AL205R AL205B to recognize the USB thumb drive and start the firmware update The Bit Sync LED flashes rapidly whi...

Страница 14: ...e Password as configured in the AL205R AL205B FIGURE 4 1 AL205B firmware upload After successfully connecting to the AL205R AL205B upload the firmware file to the root folder of the device After the u...

Страница 15: ...t may be helpful for troubleshooting 5 2 Local download USB flash drive or microSD card When a USB flash drive is inserted in the USB port the AL205R AL205B automatically copies system logs onto the U...

Страница 16: ...k functionality but the AL205R AL205B can connect directly to a computer serial port NOTE LoggerNet Admin software is required to use LoggerNet ALERT2 Administration Utility 1 Start ALERT2 Administrat...

Страница 17: ...name and select TCP IP for the Link Type Click OK b In the TCP Callback Port field enter an open TCP IP port number TIP Remote access to this port may require opening a port in the server firewall Co...

Страница 18: ...using a callback TCP IP connection 6 If using a serial port do the following a Type a base name and select Serial Port for the Link Type Click OK b Select the Serial Port to which the AL205R AL205B i...

Страница 19: ...n to access the configuration settings 9 If using TCP IP do the following a Click the IP Settings tab b In the ASCII IP Forward field type the DNS or IP address of the LoggerNet Admin server followed...

Страница 20: ...the AL205R AL205B 10 If using a serial port do the following a Click the Serial Port Settings tab b For the serial port that the AL205R AL205B is connected set the Output Mode to ASCII and the Baud R...

Страница 21: ...elp for more information 6 Port descriptions The AL205R AL205B has three serial ports with configurable functionality The serial ports operate at RS 232 line levels Table 6 1 Serial port pinout from l...

Страница 22: ...000 not enabled by default Table 6 3 p 20 provides the pinout for the RX port Table 6 4 p 21 provides the pinout for the TX port NOTE The RX radio connector can handle a maximum of 1 5 A of continuous...

Страница 23: ...acing the AL205R AL205B Pin number Label Pin function 1 AO RF data output from the AL205R AL205B 2 G Ground 3 PTT Push to talk 4 12V 12 V power provided by the AL205R AL205B 5 CS Channel select AL205R...

Страница 24: ...ther manufacturer s products that are resold by Campbell Scientific are warranted only to the limits extended by the original manufacturer Refer to www campbellsci com terms warranty for more informat...

Страница 25: ...ion or Repair Reference number contact your CAMPBELL SCIENTIFIC regional office Please write the issued number clearly on the outside of the shipping container and ship as directed For all returns the...

Страница 26: ...rohibit climbing by other persons Take reasonable precautions to secure tripod and tower sites from trespassers l Use only manufacturer recommended parts materials and tools Utility and Electrical l Y...

Страница 27: ...nfo campbellsci fr www campbellsci fr Germany Location Phone Email Website Bremen Germany 49 0 421 460974 0 info campbellsci de www campbellsci de India Location Phone Email Website New Delhi DL India...

Отзывы: