61302V Barometric Pressure Sensor 

Multipliers to convert the hPa reading to different units 

 
 
 

For

Multiply by

kPa 0.1 

Mm of Hg 

0.75006 

In of Hg 

0.02953 

mBar 1.0 

PSI 0.0145 

 
 
 
 
 
 
 
 
The weather service and most airports correct the pressure 
reading at a particular station to what it would be if that station 
were located at sea level. This is done so that weather forecasters 
can obtain a clearer picture of what is happening in the 
atmosphere.   
 
The following equation is used in the program examples to 
determine the pressure correction factor.  This can then be added 
to the measured value. Because barometric pressure is greatly 
affected by elevation, the elevation of the site used in the equation 
should be as accurate as possible.  Where Elevation is in meters 
above sea level. 
 
P = 1013.25 [ 1 – ( 1 – Elevation/44307.69231 ) ^5.253283 ] 
 

5.1. 

Serial Output Program Example 

 
The program example illustrates the polling of a 61302V in its 
default configuration using a CR1000 datalogger.  
Communications will occur on serial port COM1.  The sensor is 
powered for one minute before the top of each hour, at which time 
the sensor is polled for a measurement.  The program also 
includes the option to correct for elevation.  The measurement 
units in the example are hPa. 

 

'CR1000 Series Datalogger 
 
' Jumper P1 will need to be set to Polled mode (jumpers B & C in place) 
' Jumper P3 will need to be set to "Fixed" modes (as per Appendix A) 
' Jumper P4 will need to be set to Serial mode (A position) 
' Baud 9600, 8 bit, 1 stop, no parity. 
 
'61302V Wiring 
 
'Red - CR1000 SW12V (Unit can be powered continuously if power consumption is not an 
issue) 
'Black - CR1000 G 
'Blue - CR1000 C1 (Tx) 
'White - CR1000 C2 (Rx) 

 

9

Содержание 61302V

Страница 1: ...INSTRUCTION MANUAL 61302V Barometric Pressure Sensor September 2009 Copyright 2009 Campbell Scientific Canada Corp...

Страница 2: ...glect accidents of nature or shipping damage This warranty is in lieu of all other warranties expressed or implied including warranties of merchantability or fitness for a particular purpose CSC is no...

Страница 3: ...3 3 2 1 Serial Output Wiring 4 3 2 2 Analog Output Wiring 4 4 SENSOR OPERATION 4 4 1 Serial Output 5 4 2 Analog Output 6 4 3 Software Commands 7 5 PROGRAMMING 8 5 1 Serial Output Program Example 9 5 2...

Страница 4: ...serial I O options include continuous or polled ASCII text output or NMEA marine output The sensor is supplied with the following standard configuration RS232 serial output TTL levels Polled mode 9600...

Страница 5: ...of Campbell Scientifics white fibreglass enclosures If the 61302V is to being installed in an enclosure a QDP HYDRO VENT should used Otherwise the included poly tube can be used to vent the sensor to...

Страница 6: ...Ended Channel Single Ended Channel Blue Optional Trigger No Connection or SW12V No Connection or SW12V Green Reference AG Red PWR SW12V or 12V SW12V or 12V Jumper b w Ctrl Port SW12 CTRL Black PWR AG...

Страница 7: ...can be used with TTL RS 232 signal levels P4 A B configures the terminal connections for RS 232 Both jumpers must be in position A for RS 232 serial communication Format Description Continuous ASCII...

Страница 8: ...meters The standard setting maps 0 to 5000 mV with 500 to 1100 hPa For example hPa 0 12 mV 500 Effective range is 600 hPa Analog signal resolution is 12 bit resolution 600 212 Therefore the effective...

Страница 9: ...ile paused COMMAND mode is indicated by a prompt All commands must be terminated with a carriage return CR ASCII 13 Use to list available commands as shown below YOUNG 61300 SERIES BAROMETER V2 03 CMD...

Страница 10: ...ely after the command is received Therefore in order to continue communicating the serial communication program that issued the command must also change to match CMD300 n causes the sensor to pause or...

Страница 11: ...ssible Where Elevation is in meters above sea level P 1013 25 1 1 Elevation 44307 69231 5 253283 5 1 Serial Output Program Example The program example illustrates the polling of a 61302V in its defaul...

Страница 12: ...Open Com1 9600 0 0 1000 Scan 5 Sec 0 0 Switch 61302V Sensor On 1 minute before the top of the hour If TimeIntoInterval 59 60 Min Then SW12 1 Poll 61302V Sensor at the top of the hour If TimeIntoInterv...

Страница 13: ...set to 5VDC mode A position Jumper P4 will need to be set to Voltage mode B position 61302V Wiring Red CR1000 12V power continuously for testing Black CR1000 AG White CR1000 SE1 Blue No Connection Gre...

Страница 14: ...lling SW12 0 Correct Barometric Pressure value for elevation above sea level elevation must be in meters Elevation_Correction 1013 25 1 1 Elevation 44307 69231 5 253283 BP_61302V_Analog BP_61302V_Anal...

Страница 15: ...61302V Barometric Pressure Sensor 13...

Страница 16: ...y 1 Connect the 61302V to the PC Please note in order to make the connection to the PC you will require either an interface board i e DB9M TERM or a 9 pin cable with connections for Tx Rx and Ground T...

Страница 17: ...mit to 1050 Use CMD 142 to set the lower measurement limit to 950 These changes must be saved to flash memory which is done by entering CMD420 5 Confirm that the changes have been saved to memory by e...

Страница 18: ...Barometric Pressure Sensor 16 6 Restart the measurement process with CMD300 2 7 Once the configuration is complete you can power the sensor down and replace all jumpers required for your specific appl...

Отзывы: