229 Heat Dissipation Matric Water Potential Sensor
11: Thermocouple Temp (DIFF) (P14)
;Read thermocouple after 1 second of heating
1: 1
Reps
2: 21
10 mV, 60 Hz Reject, Slow Range
3: 1
DIFF
Channel
4: 1
Type
T
(Copper-Constantan)
5: 2
Ref Temp (Deg. C) Loc [ Tref_C ]
6: 19
-- Loc [ T1s_1 ]
7: 1.0
Mult
8: 0.0
Offset
12: Excitation with Delay (P22)
;delay 29 more seconds
1: 1
Ex
Channel
2: 0
Delay W/Ex (0.01 sec units)
3: 2900
Delay After Ex (0.01 sec units)
4: 0000
mV
Excitation
13: Thermocouple Temp (DIFF) (P14)
;read temperature after 30 seconds heating
1: 1
Reps
2: 21
10 mV, 60 Hz Reject, Slow Range
3: 1
DIFF
Channel
4: 1
Type
T
(Copper-Constantan)
5: 2
Ref Temp (Deg. C) Loc [ Tref_C ]
6: 35
-- Loc [ T30s_1 ]
7: 1.0
Mult
8: 0.0
Offset
14: Do (P86)
;turn off CE8
1: 51
Set
Port
1
Low
15: Z=X-Y (P35)
;Calculate temperature rise
1: 35
-- X Loc [ T30s_1 ]
2: 19
-- Y Loc [ T1s_1 ]
3: 5
-- Z Loc [ dT_1 ]
16: End (P95)
;end of loop
17: Do (P86)
;Turn off AM16/32
1: 52
Set
Port
2
Low
18: End (P95)
;End of then do statement
19: If Flag/Port (P91)
1: 11
Do if Flag 1 is High
2: 10
Set Output Flag High (Flag 0)
20: Set Active Storage Area (P80)
1: 1
Final
Storage
Area
1
2:
229
Array
ID
21: Real Time (P77)
1: 1220
Year,Day,Hour/Minute
(midnight
=
2400)
18
Содержание 229