Model 092 Barometric Pressure Sensor
5. Overview
The 092 ships from Campbell Scientific pre-wired for analog output with the
following settings:
•
Analog Output Voltage: 0 – 5 VDC
•
Range: 600 – 1100 mbar
Refer to Appendix A,
Analog Output Configurations
for available analog
output options.
The 092 utilizes an on-board microcontroller and piezoresistive pressure sensor
module. This module contains an analog to digital converter, a temperature
sensor, and non-volatile memory for storage of factory determined calibration
coefficients. The microcontroller polls the sensor module once per second.
Measurements are temperature corrected, the calibration coefficients applied,
and the processed pressure measurement stored for output.
The pressure sensor element is light sensitive. Do not operate
the 092 with the top cover of the polycarbonate case off.
NOTE
All wiring terminals on the 092 are protected from static surge damage by
transzorbs and current limiting resistors.
Although the 092 is configured at Campbell Scientific for analog output,
digital output options including RS-232 and SDI-12 are available. Refer to the
MetOne
092, 594, 6633A Barometric Pressure Sensor Manual
for more
information.