59/89 Device handbook A
PLUS
with / without LED display
,
157
679-19, 04/2016
Security
The current inputs are galvanically isolated from each other
Protection class:
II (protective insulation, voltage inputs via protective impedance)
Pollution degree:
2
Protection:
IP64 (front), IP40 (housing), IP20 (terminals)
Measurement category:
CAT III, CATII (relays)
Rated voltage
power supply: 265 V AC
(versus earth):
Relays: 250 V AC
I/O’s: 30 V DC
Test voltages:
DC, 1 min., acc. IEC/EN 61010-1
7504V DC, power supply versus inputs U, I
5008V DC, power supply versus bus, I/O’s, relays
6030V DC, inputs U versus inputs I
4690V DC, inputs U after protective impedance versus bus, I/O’s, relays
7504V DC, inputs U versus relays
7504V DC, inputs I versus bus, I/O’s, relays
6030V DC, inputs I versus inputs I
3130V DC, relay versus relay, bus, I/O's
Applied regulations, standards and directives
IEC/EN 61 010-1
Safety regulations for electrical measuring, control and laboratory equipment
IEC/EN 60 688
Electrical measuring transducers for converting AC electrical variables into
analog or digital signals
DIN 40 110
AC quantities
IEC/EN 60 068-2-1/
Ambient tests
-2/-3/-6/-27:
-1 Cold, -2 Dry heat, -3 Damp heat, -6 Vibration, -27 Shock
IEC/EN 60 529
Protection type by case
IEC/EN 61 000-6-2/
Electromagnetic compatibility (EMC)
61 000-6-4:
Generic standard for industrial environment
IEC/EN 61 131-2
Programmable controllers - equipment, requirements and tests
(digital inputs/outputs 12/24V DC)
IEC/EN 61 326
Electrical equipment for measurement, control and laboratory use - EMC
requirements
IEC/EN 62 053-31
Pulse output devices for electromechanical and electronic meters (S0 output)
UL94
Tests for flammability of plastic materials for parts in devices and appliances
2011/65/EU (RoHS)
EU directive on the restriction of the use of certain hazardous substances
Warning
This is a class A product. In a domestic environment this product may cause radio interference in which
case the user may be required to take adequate measures.
This device complies with part 15 of the FCC:
Operation is subject to the following two conditions: (1) This device may not cause harmful interference,
and (2) this device must accept any interference received, including interference that may cause
undesired operation.
This Class A digital apparatus complies with Canadian ICES-0003.