Cameron NUFLO Scanner 2000 microEFM Скачать руководство пользователя страница 5

5

Calibrate Inputs Menu

From the Calibrate Inputs menu, a user can calibrate differential pressure, static pressure, and process temperature. The user can load a 

factory calibration, load a previous calibration, or enter new calibration points.  

3

  The measured value 

can be viewed with or 
without a trend graph. 
Repeat Step 2 to enter 
additional calibration 
points. 

Click on Change Task to 
select a new parameter 
or a new calibration task. 

1

  Select the task you 

want to perform. 
Then select the parameter 
you want to calibrate.

2

  Click Acquire Point to enter a new 

calibration  point and enter the applied 
calibration value at the prompt. 
Or click Load Calibration From... to load 
the data from a factory calibration or  
previous calibration. 

4

  Click Save Changes to 

save the New Calibration 
values to memory.

5

  Click Exit 

Maintenance Mode  
to resume live data 
readings. 

Maintenance Mode

When a user enters a menu in which parameter changes 

affect volume calculations, the  Scanner 2000 is placed in 

“Maintenance Mode.” This locks the input sensor readings at 

their current values. When the user exits “Maintenance Mode,” 

the inputs are unlocked.  

Содержание NUFLO Scanner 2000 microEFM

Страница 1: ...ay need to be adjusted To change the orientation of the head loosen the set screw in the MVT adapter rotate the head no more than 180 degrees and tighten the set screw Remote Mount To install the Scan...

Страница 2: ...UP B A TB1 TFM1 I R R I 2 3 4 5 6 1 BLACK RED DIGITAL OUTPUT 13 14 Leave this end of shield disconnected POWER SUPPLY 5 to 30 VDC Resistor may be included in pulse readout device Size the resistor to...

Страница 3: ...digit to the left Repeat using UP ARROW LOG and LEFT ARROW DISPLAY to enter all remaining digits Press ENTER SAVE Keypad Configuration Controls Remove the instrument cover to access the keypad Each of...

Страница 4: ...is the control center for the Scanner 2000 and provides quick and easy access to the most commonly performed tasks From this screen the user can view real time totals and flow rates input data and sy...

Страница 5: ...o perform Then select the parameter you want to calibrate 2 Click Acquire Point to enter a new calibration point and enter the applied calibration value at the prompt Or click Load Calibration From to...

Страница 6: ...od used Click Change Task to select a new maintenance task 1 Select the task you want to perform 3 Click Save Changes to save the new parameters to memory 2 Enter new parameters in the fields provided...

Страница 7: ...to specify plate parameters 9 Enter base conditions specific to the location 10 Click Change Composition to specify the gas composition Percentages must total 100 8 Specify pipe parameters 11 Click O...

Страница 8: ...screen Download Screens Part No 9A 30165024 Rev 03 To download data simply press the Download button and select the range and type of files to be downloaded Daily and interval logs can be viewed or p...

Отзывы: