CD-20SFA Sample Probe Controller 

 

Section 1—Overview

The CD-20SFA Sample Probe Controller is engineered to 

provide pulse input (flow proportional) control 

of the True-Cut C-Series sample probe. 

When a “start” pulse is received, the probe is rotated to capture and 

discharge an isokinetic sample. When installed in a horizontal line, the probe is stopped in the closed position, 

while on a vertical line the probe is stopped in the open position. 
The C21 probe collects one 1.5 cc sample per 360-degree rotation.
The C22 probe collects two 1.5 cc samples per 360-degree rotation.
A failure alarm relay provides an output to alert users to a loss of power, loss of input signal, or motor failure. 

The relay contact remains open (inactive) for as long as a sample is being taken within the alarm delay period 

configured with the control card potentiometer. Alarm delays can range from 5 seconds to 6 minutes. 
An explosion-proof housing encloses the complete assembly, and two ¾-in. NPT holes allow connection of 

the line power and input pulse (for flow-proportional sampling). A ½-in. breather plug and a ½-in. conduit 

plug are installed on opposite sides of the enclosure.
A coupling, jam nut, and hub allow the assembly to be mounted directly to the C-Series sample probe. No 

brackets or supports are required; however, they are recommended for piping with significant vibration.
The CD-20SFA controller consists of the following:
•  a motor control assembly (24VDC, 115VAC, or 230VAC) 
•  a DC gear motor with a gear ratio of 150:1
•  a cam with two high points mounted on the gear motor output shaft
•  a proximity switch mounted in the hub assembly opposite the cam
When a high point on the cam is opposite the proximity switch, the PROX (“Hold”) LED is illuminated. This 

signal indicates the sample probe is at the “stop and hold” position.

Figure 1.1—CD-20SFA controller components

Содержание CLIF MOCK CD-20SFA

Страница 1: ...CLIF MOCK CD 20SFA Sample Probe Controller User Manual Manual No 2350754 01 Rev 02 Class I Groups C and D Hazardous Locations...

Страница 2: ...d within the parameters stated in this manual and in conjunction with the ratings printed on the product serial tag Technical Support Contact Information Cameron Measurement Systems Division 14450 Joh...

Страница 3: ...5 How It Works 6 Hazardous Location Safety Compliance 6 Table 1 1 CD 20SFA Specifications 7 Section 2 Installation and Wiring 7 Section 3 Maintenance 8 Operational Test 8 Basic CD 20SFA Check Out 9 S...

Страница 4: ...ler s liability and Buyer s exclusive remedy in any case of action whether in contract tort breach of warranty or otherwise arising out of the sale or use of any products software supplies or material...

Страница 5: ...control card potentiometer Alarm delays can range from 5 seconds to 6 minutes An explosion proof housing encloses the complete assembly and two in NPT holes allow connection of the line power and inpu...

Страница 6: ...rols the speed of the motor at about 12 rpm and current feedback limits motor current to 3 amperes The pulse input be closed for at least 5 msec to produce a start pulse An electronic square wave inpu...

Страница 7: ...bly earthed and installed by qualified personnel in accordance with the prevailing local electrical wiring regulations and safety standards The CD 20SFA controller mounts directly to the C Series samp...

Страница 8: ...failure alarm relay contact to 1TB 3 and 1TB 4 The contact will remain open for as long as sampling occurs within the alarm delay period This contact is rated for 90 VA resistive and 30 VA inductive...

Страница 9: ...to 1TB 5 and 1TB 6 2 Mount the CD 20SFA to the sample probe and hand tighten the DC 2 coupling 3 Verify that the product line is full and pressurized 4 Reconnect input power to the CD 20SFA 5 Hold a...

Страница 10: ...D1 a 33V crowbar zener diode pro vides over voltage protection for the rest of the circuit Diode D2 prevents current flow if VDC power is connected to the common terminal and a ground is connected to...

Страница 11: ...mity switch is not sensing the high points of the cam Check proximity switch and cam alignment CD 20SFA control failure power supply board Return to factory for replacement of CD 20SFA motor control a...

Страница 12: ...R U S S I A 603 5569 0501 ms kl c a m com...

Отзывы: