Aidite |
24
4.14 Sleep mode interface
The user can manually or automatically
enable the furnace to enter the sleep
state. The platform rises to close the
furnace. Keep the temperature constant
at 100° C. Click anywhere on the screen
to exit sleep mode.
There are two ways to enter the sleep
mode:
I. Click “ ” on the main interface.
II. After four continuous hours of
no operation activities on the main
interface.
4.15 Furnace cleaning
Click Furnace Clean in the function
interface, the tray will be lowered
to the lowest end. Put the crucible
with activated carbon on the firing
platform. Click Confirm, and wait till
the temperature is between 190° C
and 210° . The tray automatically rises
until the furnace door is closed and the
furnace cleaning starts, which lasts for
30 minutes.