Calibration
Menu Definitions
:
C
1
Zero All Raw counts, (pitch and roll if in angle mode) will be displayed.
When ZERO is pressed an analog
z
ero is done and all
calibration span points will be cleared.
If in angle mode the pitch and roll offsets will also be zeroed.
C
2
Zero
Zeroed Raw counts, (pitch and roll if in angle mode will be
displayed
)
. When ZERO is pressed an analog zero is done
and all calibration span points will NOT be cleared.
If in angle mode the pitch and roll offsets will also be zeroed.
C
3
Span Point The last Calibration weight will be displayed then the
actual weight on the scale will be displayed.
If you do not wish to change the span point, press the
TARE key to exit without making any changes.
If the displayed weight does not match the known test
weight, press the ZERO key to enter the correct weight.
Use the steps d
e
scribed in section
2
.4 for numeric entry.
When the weight is correct press the Print key to enter
the new value.
C4 to C7
Span points C4 to C7 are for linearity correction, they can be used in
order and in any quantity or not at all if no correction is
necessary.
You may also return to C4 to C7 later and add a new
correction point without
a
ffecting any original calibration
points.
The last calibration weight will be displayed then the actual
weight on the scale will be displayed. If no calibration weight
has been entered at this span point
"notset"
will be displayed
then the actual weight on the scale is displayed. If the
displayed weight does not match the known test
weight,
follow the steps d
e
scribed for C3 Span point on
adjusting the
weight and entering the value.
END
Exits back to the main Menu.
3
.0 CALIBRATION PROCEDURES
CAMBRIDGE SCALE WORKS, INC.
10
MANUAL P/N 5999-1014-01