background image

Содержание CSW-10AT

Страница 1: ......

Страница 2: ...nu Definitions 7 8 2 4 Numeric Entries 9 2 5 Set Up Parameters 9 3 0 Calibration Procedures 10 3 1 Calibration Menu Definitions 10 3 2 Calibration 11 3 3 Linearity Correction 11 4 0 Communications Setup 12 4 1 Communications Menu Definitions 12 13 5 0 Testing Procedure 14 5 1 Testing Menu Definitions 14 6 0 Warranty 15 7 0 Assistance 16 CAMBRIDGE SCALE WORKS INC 2 MANUAL P N 5999 1014 01 ...

Страница 3: ...osition Connect the Transformer to y your 110VAC outlet Next connect the other End of the cable to the Rear of the Meter for Power Meter Connections Communications port Signal connection PIN DESCRIPTION PIN DESCRIPTION 2 RXD 1 EXC 3 TXD 2 SIG 4 5V 5 SHIELD LOAD CELL 6 EXC 7 SIG 5 SGND CAMBRIDGE SCALE WORKS INC 3 MANUAL P N 5999 1014 01 ...

Страница 4: ... the Gross weight value into the Tare display and switches to the Net display mode If the Tare key is pressed and held for 5 seconds the current Tare value will be displayed lb kg Toggles the display between pounds and kilograms PRINT Outputs the displayed weight data to the RS 232 Port Note All keys are disabled when the scale is in motion or overload 1 2 Error Messages ScnEg When the weight is m...

Страница 5: ... ZERO SHOW RAW COUNTS THEN 0 SHOW RAW COUNTS THEN 0 SHOW ACTUAL WEIGHT C3 SPAN POINT C4 SPAN POINT C5 SPAN POINT C6 SPAN POINT C7 SPAN POINT END R1 BAUD RATE R2 Output Format R3 Output Type R4 Receiver mode 1200 8 N 1 19200 8 N 1 WT ONLY G N GTN Key Polled Continuous Disabled Rec Remote END T1 VERSION T2 DISPLAY LIGHT SEGMENTS SOFTWARE VERSION T3 BUTTONS b1 b4 b5 EXITS T4 A TO D A D RAW COUNTS T5 ...

Страница 6: ... IS USED TO NAVIGATE RIGHT PRINT KEY IS USED TO ENTER DATA AND RETURNS 2 1 SOFTWARE NAVIGATION FLOWCHART CONTINUED S9 Motion Display Normal Blank Dashes Normal S10 ZERO BAND 1 99 DIVISIONS 2 S11 ZERO DELAY 0 99 UPDATES 2 1 99 DIVISIONS 1 1 99 1 4secinc 0 lb kg lb kg lb kg S12 Zero Tracking S13 Tracking DELAY S14 Display lb kg S15 SCALE ID S16 BRIGHTNESS S17 SLEEP MODE END 1 99 1 ID IN TX 1 15 15 B...

Страница 7: ...1 9 of scale capacity S6 Filter 0 7 3 0 is the fastest response or least filtering and 7 is the slowest response or most filtering S7 Motion Band 1 to 99 divisions 2 the weight display must be stable within the selected number of division for the motion indicator to be turned off S8 Motion Delay 0 to 99 updates 4 the weight display must be within the motion band for the selected number of updates ...

Страница 8: ...ed 0 disables zero tracking S14 Lb Kg Lb KG Allows the indicator to be switched between pounds and kilograms by pressing the lb kg key LB This sets the display to Pounds only Kg This sets the display to Kilograms only S15 Scale ID 1 to 99 scale ID used in RF link output S16 Brightness 0 to 15 adjusts the LED display intensity 15 is the brightest S17 Sleep Mode 0 to 30 minutes The display will turn...

Страница 9: ...changes are made to the Scale or Calibration Parameters With the display indicating ScAlE press the GRS NET key to move down This allows the operator to change any of the scale Parameters S1 thru S17 For example press the GRS NET key to move down until S2 is displayed S2 is used to set the Capacity of the Scale Press the lb kg key to move right in the S2 Parameter The current capacity will be disp...

Страница 10: ...own test weight press the ZERO key to enter the correct weight Use the steps described in section 2 4 for numeric entry When the weight is correct press the Print key to enter the new value C4 to C7 Span points C4 to C7 are for linearity correction they can be used in order and in any quantity or not at all if no correction is necessary You may also return to C4 to C7 later and add a new correctio...

Страница 11: ...ey to move left the next digit will flash Repeat the steps until the correct weight is entered Press the PRINT key to record the data The display will return to C3 Press the GRS NET key until End is displayed then Press PRINT Calibration is now complete 3 3 Linearity Correction If Linearity Correction is needed Press the GRS NET key from the calibration menu to move down to C4 Press the lb kg key ...

Страница 12: ...GR Sp Six ASCII Charactes indicated weight Sp lb or kg CR LF 14 bytes total output R3 Output Type 0 Output on command standard print Output as selected by R2 Output Format If Q is received on the serial port the scale will output the same as if the PRINT key were Pressed The same holds true for Z Zero U lb kg D GRS NET T TARE 1 Slave Display Output numeric only continuous Stx Six ASCII Characters ...

Страница 13: ...laying data as it appears on the scale All keys are disabled except the PRINT key 2 Remote control Receives R3 3 RF link output string displaying data as it appears on the scale and allows full control of all scale meter functions END Exits back to the main Menu indicates Factory Default setting CAMBRIDGE SCALE WORKS INC 13 MANUAL P N 5999 1014 01 ...

Страница 14: ...counts When in angle mode Down GRS NET will cycle Pitch Roll and Raw counts T5 Serial Serial communications can be verified by connecting pins 2 and 3 on the serial port A single character will be echoed and pass or fail will be displayed T6 Setup data Setup data will be sent out on the printer port T7 Default Resets the meter back to factory defaults clearing all calibration and setup data r you ...

Страница 15: ...of its products beyond that stated above Warranty replacement parts and or repair services are performed at the factory in Cumberland Maryland or by an authorized Service Group approved by CAMBRIDGE Warranty does not include travel expense if a factory technician is requested to perform repairs at a location other than the factory It s the user s responsibility to follow the proper set up calibrat...

Страница 16: ...sistance with your Model CSW 10AT Meter please contact us at CAMBRIDGE SCALE WORKS INC 115 West Mary Street Cumberland MD 21502 Phone 301 724 4082 Fax 301 724 4964 7 0 Assistance CAMBRIDGE SCALE WORKS INC 16 MANUAL P N 5999 1014 01 ...

Отзывы: