 1.3 Scale Menu Definitions Continued:

 
      

S11

    Zero Delay           0 to 99 updates (4) the weight display must be within  

                                             the zero band for the selected number of updates to
                                             be considered Zero. 
                                                                                                                           
      

S12   

Zero Tracking       1 to 99 divisions (2) the number of graduations 

                                             allowed to be Automatically Zeroed off.
 
      

S13   

Tracking Delay     0 to 99 [15second increments] (10) the amount of 

                                             time the display must be within the allowed
                                             graduations before it will be automatically Zeroed.
     
      

S14    

Lb/Kg   (Lb/KG): Allows the indicator to be switched between pounds

                                             and kilograms by pressing the lb/kg key.
                                      LB: This sets the display to Pounds only.
                                      Kg: This sets the display to Kilograms only.
 
      

S15    

Scale ID               (1) to 99 scale ID used in RF link output.

 
      

S16    

Brightness            0 to (15) adjusts the LED display intensity 15 is

                                             the brightest.
 
      

S17    

Sleep Mode          (0) to 30 minutes. The display will turn off after the   

                                              the set amount of time elapses with no scale activity.
   
      

END

                                Exits back to the main Menu.

                                               
 
 
 
 
 
 
( ) indicates Factory Set defaults.                                               

MANUAL P/N: 5999-1021-01 (8/11)

9

Содержание CSW-10AT-CS

Страница 1: ...ODEL CSW 10AT CS METER SET UP OPERATION CAMBRIDGE SCALE WORKS INC PH 800 292 7640 5011 HORSESHOE PIKE FX 610 273 3612 PO BOX 670 HONEY BROOK PA 19344 www cambridgescale com MANUAL P N 5999 1021 01 8 11 1 ...

Страница 2: ... 4 Numeric Entries 10 1 5 Set Up Parameters 10 2 0 Calibration Procedures 11 2 1 Calibration Menu Definitions 11 2 2 Calibration 12 2 3 Linearity Correction 12 3 0 Communications Setup 13 3 1 Communications Menu Definitions 13 14 4 0 Testing Procedure 15 4 1 Testing Menu Definitions 15 5 0 Warnings 16 6 0 Warranty 17 7 0 Assistance 18 MANUAL P N 5999 1021 01 8 11 2 ...

Страница 3: ...witch on the Rear of the meter is in the OFF position Insert the battery in the battery holder with the contacts pointing down then slide down to lock into position Meter Connections ON OFF SWITCH CHARGE INDICATING LED CHARGE INDICATING LED MANUAL P N 5999 1021 01 8 11 3 ...

Страница 4: ...nters the Gross weight value into the Tare display and switches to the Net display mode If the Tare key is pressed and held for 5 seconds the current Tare value will be displayed lb kg Toggles the display between pounds and kilograms PRINT Outputs the displayed weight data to the RS 232 Port Note All keys are disabled when the scale is in motion or overload Error Messages SCnEg When the weight is ...

Страница 5: ...the power adapter to a 110VAC outlet Next connect the other end of the adapter to the power jack on the rear of the meter The charge indicating led will be orange while the battery is being charged and turned off when the battery is fully charged Turn the on off switch to the ON postion and the meter will be fully operational as the battery is being charged OPERATION CONTINUED CHARGE INDICATING LE...

Страница 6: ...l Points C2 ZERO SHOW RAW COUNTS THEN 0 SHOW RAW COUNTS THEN 0 SHOW ACTUAL WEIGHT C3 SPAN POINT C4 SPAN POINT C5 SPAN POINT C6 SPAN POINT C7 SPAN POINT END R1 BAUD RATE R2 Output Format R3 Output Type R4 Receiver mode 1200 8 N 1 19200 8 N 1 WT ONLY G N GTN Key Polled Continuous Disabled Rec Remote END T1 VERSION T2 DISPLAY LIGHT SEGMENTS SOFTWARE VERSION T3 BUTTONS b1 b4 b5 EXITS T4 A TO D A D RAW...

Страница 7: ...ED TO NAVIGATE LEFT lb kg KEY IS USED TO NAVIGATE RIGHT PRINT KEY IS USED TO ENTER DATA AND RETURNS 1 1 SOFTWARE NAVIGATION FLOWCHART CONTINUED S9 Motion Display Normal Blank Dashes Normal S10 ZERO BAND 1 99 DIVISIONS 2 S11 ZERO DELAY 0 99 UPDATES 2 1 99 DIVISIONS 1 1 99 1 4secinc 0 lb kg lb kg lb kg S12 Zero Tracking S13 Tracking DELAY S14 Display lb kg S15 SCALE ID S16 BRIGHTNESS S17 SLEEP MODE ...

Страница 8: ...mit 100 or 1 9 of scale capacity S6 Filter 0 7 3 0 is the fastest response or least filtering and 7 is the slowest response or most filtering S7 Motion Band 1 to 99 updates 2 the weight display must be stable within the selected number of updates for the motion indicator to be turned off S8 Motion Delay 0 to 99 updates 4 the weight display must be within the motion band for the selected number of ...

Страница 9: ... graduations before it will be automatically Zeroed S14 Lb Kg Lb KG Allows the indicator to be switched between pounds and kilograms by pressing the lb kg key LB This sets the display to Pounds only Kg This sets the display to Kilograms only S15 Scale ID 1 to 99 scale ID used in RF link output S16 Brightness 0 to 15 adjusts the LED display intensity 15 is the brightest S17 Sleep Mode 0 to 30 minut...

Страница 10: ...changes are made to the Scale or Calibration Parameters With the display indicating ScAlE press the GRS NET key to move down This allows the operator to change any of the scale Parameters S1 thru S17 For example press the GRS NET key to move down until S2 is displayed S2 is used to set the Capacity of the Scale Press the lb kg key to move right in the S2 Parameter The current capacity will be disp...

Страница 11: ...tch the known test weight press the ZERO key to enter the correct weight Use the steps discribed in section 1 4 for numeric entry When the weight is correct press the Print key to enter the new value C4 to C7 Span points C4 to C7 are for linearity correction they can be used in order and in any quantity or not at all if no correction is necessary You may also return to C4 to C7 later and add a new...

Страница 12: ... Repeat these steps until the correct weight is entered Press the PRINT key to Enter the data The display will return to C3 Press the GRS NET key until End is displayed then Press PRINT to return to the operation mode Calibration is now complete 2 3 Linearity Correction If Linearity Correction is needed Press the GRS NET key from the calibration menu to move down to C4 Press the lb kg key to move ...

Страница 13: ...Gross only GR Sp Six ASCII Charactes indicated weight Sp lb or kg CR LF 14 bytes total output R3 Output Type 0 Output on command standard print Output as selected by R2 Output Format If Q is received on the serial port the scale will output the same as if the PRINT key were Pressed The same holds true for Z Zero U lb kg D GRS NET T TARE 1 Slave Display Output numeric only continuous Stx Six ASCII ...

Страница 14: ...ying data as it appears on the scale All keys are disabled except the PRINT key 2 Remote control Receives R3 3 RF link output string displaying data as it appears on the scale and allows full control of all scale meter functions END Exits back to the main Menu indicates Factory Default setting 3 0 COMMUNICATIONS SETUP MANUAL P N 5999 1021 01 8 11 14 ...

Страница 15: ...ay 25 000 counts When in angle mode Down GRS NET will cycle Pitch Roll and Raw counts T5 Serial Serial communications can be verified by connecting pins 2 and 3 on the serial port A single character will be echoed and pass or fail will be displayed T6 Setup data Setup data will be sent out on the printer port T7 Default Resets the meter back to factory defaults clearing all calibration and setup d...

Страница 16: ...orn beyond the limits shown in Figure 1 Remove from service any hook with a crack nick or gouge Hooks with a crack nick or gouge shall be repaired by grinding lengthwise following the contour of the hook provided that the reduced dimension is within the limits shown in Figure 1 Never repair alter rework or reshape a hook by welding heating burning or bending Never side load back load or tip load a...

Страница 17: ...ection with the sales of its products beyond that stated above Warranty replacement parts and or repair services are performed at the factory in Cumberland Maryland or by an authorized Service Group approved by CAMBRIDGE Warranty does not include travel expense if a factory technician is requested to perform repairs at a location other than the factory It the user s responsibility to follow the pr...

Страница 18: ...ou require assistance with you Model CSW 10AT CS Meter please contact us at CAMBRIDGE SCALE WORKS INC 11500 Bedford Road Cumberland MD 21502 301 724 4082 Fax 301 724 4964 7 0 Assistance MANUAL P N 5999 1021 01 8 11 18 ...